EP-WORKMATE¿ SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 56 CHANNEL
Report
- Report Number
- 2184149-2018-00068
- Event Type
- Injury
- Date Received
- June 4, 2018
- Date of Event
- May 7, 2018
- Report Date
- August 21, 2018
- Manufacturer
- ST. JUDE MEDICAL, INC.
- Product Code
- DQK
- PMA / PMN Number
- K132073
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
THE USER RECEIVED A BURN AT THE LOCATION OF THE SHOCK. ONE EP-WORKMATE SCU (SIGNAL CONDITIONING UNIT) WAS RECEIVED FOR EVALUATION. CONTINUITY TESTING OF PINS 35 AND 36 SHOWED NO SHORTS. USING A TEST STANDARD SCSI CABLE, A BASIC SIGNAL ACQUISITION/QUALITY TEST WHICH INCLUDED THE BASELINE, AMPLITUDE, AND SWITCHING WERE PERFORMED AND CONFIRMED THAT THE RETURNED AMPLIFIER PERFORMED WITHIN THE FACTORY SPECIFICATIONS. AN SCU SAFETY TEST WAS RUN AND THE UNIT PASSED. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE REVIEW DETERMINED THE PROCESS WAS PERFORMED AND COMPLETED IN ACCORDANCE WITH ABBOTT SPECIFICATIONS AND PROCEDURES. THE REPORTED COMPLAINT THAT THE NURSE RECEIVED AN ELECTRICAL SHOCK WHEN THE SCSI CABLE WAS TOUCHED COULD NOT BE REPRODUCED DURING THE EVALUATION.
DURING AN ATRIOVENTRICULAR NODE ABLATION PROCEDURE A USER SHOCK OCCURRED. DURING THE PROCEDURE THE SIGNAL FROM THE ABLATION CATHETER WAS LOST. THE CATHETER, CABLING AND PINS LOCATIONS WERE CHANGED BUT THE ISSUE WAS NOT RESOLVED. IT WAS SUGGESTED TO CHANGE THE SCSI CONNECTOR AND WHEN THE NURSE TOUCHED THE CONNECTOR SHE RECEIVED AN ELECTRIC SHOCK. THE PROCEDURE WAS STOPPED DUE TO THIS ISSUE. THERE WERE NO ADVERSE PATIENT OR USER CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 408184 | EP-WORKMATE¿ SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 56 CHANNEL | PROGRAMMABLE DIAGNOSTIC COMPUTER | DQK | ST. JUDE MEDICAL, INC. | 09-1462-0056 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |