FDA Adverse Event Death Summary report: N

BD BACTEC¿ PEDS PLUS¿/ F CULTURE VIALS

MDR report key: 7562714 · Received June 1, 2018

Report

Report Number
2647876-2018-00017
Event Type
Death
Date Received
June 1, 2018
Date of Event
April 10, 2018
Report Date
May 30, 2018
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904421946
PMA / PMN Number
K954927
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. INVESTIGATION: BD DIAGNOSTIC SYSTEMS (BDDS) QUALITY INVESTIGATED THE COMPLAINT OF A POTENTIAL FALSE POSITIVE BIOFIRE TEST RESULT WHEN USED WITH BD BACTEC¿ PEDS PLUS¿/ F CULTURE VIAL. THE BIOFIRE BCID IDENTIFIED PROTEUS IN A POSITIVE PATIENT BLOOD CULTURE BOTTLE WITH THE RESULTS BEING REPORTED OUT AFTER THE PATIENT EXPIRED. THE PATIENT WAS 3 MONTHS IN AGE AND HAD BEEN IN THE NICU FOR AN INFECTION SINCE BIRTH. THE CAREGIVERS WERE AWARE THAT THE PATIENT HAD AN E.COLI INFECTION PRIOR TO RECEIVING THE BIOFIRE RESULTS. THE PATIENT WAS TREATED WITH GENTAMICIN AND VANCOMYCIN. PROTEUS DID NOT GROW IN CULTURE NOR WAS IT SEEN ON GRAM STAIN. ACCORDING TO THE HOSPITAL THE PATIENT¿S DEATH WAS NOT DUE TO ANY CHANGES IN TREATMENT. BDDS INVESTIGATION FINDINGS TO DATE ARE AS FOLLOWS: WE HAVE TESTED MORE THAN 400 SAMPLES FOR BIOBURDEN AND/OR MOLECULAR TESTING. NO VIABLE PROTEUS ORGANISMS HAVE BEEN RECOVERED IN ANY SAMPLE INCLUDING RAW MATERIALS, IN-PROCESS WATER, THE DEIONIZED (DI) WATER SYSTEM, AND THE ENVIRONMENTAL SAMPLES THROUGHOUT THE MANUFACTURING PROCESS. NO PROTEUS DNA HAS BEEN DETECTED BY MOLECULAR TESTING WITH BIOFIRE FILMARRAY BCID ASSAY IN THE WATER SUPPLY, THE DI WATER SYSTEM AND THE ENVIRONMENTAL SAMPLES THROUGHOUT THE MANUFACTURING PROCESS. CONSISTENT WITH CUSTOMER REPORTS, SOME LOTS OF FINAL PRODUCT TESTED POSITIVE FOR PROTEUS SP WITH BIOFIRE FILMARRAY BCID ASSAY. THERE HAVE BEEN NO RECENT CHANGES TO RAW MATERIALS, DESIGN, MANUFACTURING PROCESS, OR DISTRIBUTION PRACTICES. THE REMAINDER OF BDDS INVESTIGATION WILL FOCUS ON THE FOLLOWING AREAS: ADDITIONAL TESTING OF THE DEHYDRATED MEDIA AND THE RAW MATERIAL COMPONENTS. CROSS-REACTIVITY EVALUATION OF PURIFIED CULTURES OF ENVIRONMENTAL ORGANISMS FOUND DURING BIOBURDEN TESTING. TESTING OF ADDITIONAL IN-PROCESS SAMPLES. FURTHER ACTIONS ARE TO BE DETERMINED BASED ON THE REMAINING INVESTIGATION RESULTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT BIOFIRE IDENTIFIED PROTEUS IN A BD BACTEC¿ PEDS PLUS¿/ F CULTURE VIAL WITH RESULTS BEING REPORTED OUT AFTER THE PATIENT EXPIRED. THE PATIENT WAS THREE MONTHS IN AGE AND HAD BEEN IN THE NICU FOR THE INFECTION SINCE BIRTH. THE CAREGIVERS WERE AWARE THAT THE PATIENT HAD E.COLI PRIOR TO RECEIVING THE BIOFIRE RESULTS. THE PATIENT WAS TREATED WITH GENTAMICIN AND VANCOMYCIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
405147 BD BACTEC¿ PEDS PLUS¿/ F CULTURE VIALS CULTURE VIAL MDB BECTON DICKINSON CARIBE LTD. 8015973 00382904421946

Patients

Seq Age Sex Outcome Treatment
1 3 MO Death