FDA Adverse Event Malfunction Summary report: N

VERSAFITCUP CC TERMINAL CUP IMPACTOR FOR METAL BACK CC

MDR report key: 7560669 · Received June 1, 2018

Report

Report Number
3005180920-2018-00393
Event Type
Malfunction
Date Received
June 1, 2018
Date of Event
May 3, 2018
Report Date
June 1, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 01 JUNE 2018: LOT 1213128: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 05 FEBRUARY 2013. NO ANOMALIES FOUND RELATED TO THE ISSUE. A SIMILAR EVENT ON THE SAME LOT REGISTERED UP TO NOW. VISUAL INSPECTION PERFORMED ON 29 MAY 2018 BY R&D PRODUCT MANAGER: DURING THE ANALYSIS IT IS EVALUATED THAT THE TERMINAL IS BLOCKED INSIDE THE CUP. WE HAVE DISLOCATED THE TERMINAL FROM THE CUP USING THE SCREWDRIVER REF (B)(4). LOOKING AT THE THREADED PART OF THE TERMINAL WE EVALUATED THAT THE TIP PART IS LITTLE BIT DEFORMED. IN ADDITION THE BALINIT-C COATING IS SCRATCHED. ALSO ON THE CUP THERE ARE SOME RESIDUAL OF BALINIT-C. IT SEEM THAT THE SURGEON NOT SCREWED CORRECTLY THE TERMINAL IN THE CUP BEFORE IMPACTION AND THIS CAUSED THE DEFORMATION OF THE THREADED TIP TOGETHER WITH THE SCRATCHED OF THE BALINIT-C COATING.

Description of Event or Problem · 1

THE TERMINAL CUP IMPACTOR STUCK INTO THE CUP AND CAN'T BE DISCONNECTED. THE SURGEON USED ANOTHER CUP AND THE LONG STRAIGHT MULTIFUNCTION HANDLE TO COMPLETE THE SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
403216 VERSAFITCUP CC TERMINAL CUP IMPACTOR FOR METAL BACK CC SURGICAL INSTRUMENT FOR HIP LXH MEDACTA INTERNATIONAL SA 1213128

Patients

Seq Age Sex Outcome Treatment
1 Other