VERSAFITCUP CC TERMINAL CUP IMPACTOR FOR METAL BACK CC
Report
- Report Number
- 3005180920-2018-00393
- Event Type
- Malfunction
- Date Received
- June 1, 2018
- Date of Event
- May 3, 2018
- Report Date
- June 1, 2018
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
BATCH REVIEW PERFORMED ON 01 JUNE 2018: LOT 1213128: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 05 FEBRUARY 2013. NO ANOMALIES FOUND RELATED TO THE ISSUE. A SIMILAR EVENT ON THE SAME LOT REGISTERED UP TO NOW. VISUAL INSPECTION PERFORMED ON 29 MAY 2018 BY R&D PRODUCT MANAGER: DURING THE ANALYSIS IT IS EVALUATED THAT THE TERMINAL IS BLOCKED INSIDE THE CUP. WE HAVE DISLOCATED THE TERMINAL FROM THE CUP USING THE SCREWDRIVER REF (B)(4). LOOKING AT THE THREADED PART OF THE TERMINAL WE EVALUATED THAT THE TIP PART IS LITTLE BIT DEFORMED. IN ADDITION THE BALINIT-C COATING IS SCRATCHED. ALSO ON THE CUP THERE ARE SOME RESIDUAL OF BALINIT-C. IT SEEM THAT THE SURGEON NOT SCREWED CORRECTLY THE TERMINAL IN THE CUP BEFORE IMPACTION AND THIS CAUSED THE DEFORMATION OF THE THREADED TIP TOGETHER WITH THE SCRATCHED OF THE BALINIT-C COATING.
THE TERMINAL CUP IMPACTOR STUCK INTO THE CUP AND CAN'T BE DISCONNECTED. THE SURGEON USED ANOTHER CUP AND THE LONG STRAIGHT MULTIFUNCTION HANDLE TO COMPLETE THE SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 403216 | VERSAFITCUP CC TERMINAL CUP IMPACTOR FOR METAL BACK CC | SURGICAL INSTRUMENT FOR HIP | LXH | MEDACTA INTERNATIONAL SA | 1213128 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |