FDA Adverse Event Injury Summary report: N

"2.4MM" SYSTEM HIGH TORQUE (HT) CROSS-DRIVE SCREW

MDR report key: 7560462 · Received June 1, 2018

Report

Report Number
0001032347-2018-00334
Event Type
Injury
Date Received
June 1, 2018
Date of Event
November 26, 2018
Report Date
April 29, 2019
Manufacturer
BIOMET MICROFIXATION
Product Code
HWC
PMA / PMN Number
PK001238
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. REPORTED EVENT WAS CONFIRMED BY EVIDENCE THAT THE REVISION WAS PERFORMED. THESE PARTS WERE REMOVED IN A REVISION DUE TO INFECTION; THEREFORE THE COMPLAINT IS CONSIDERED CONFIRMED. NO X-RAYS, SCANS, PICTURES, OR PHYSICIANS REPORTS WERE PROVIDED. NO DETAILS WERE PROVIDED AS TO THE CAUSE OF THE INFECTION. ALL TMJPMS ARE DISTRIBUTED STERILE. THE STERILE CERTIFICATIONS FOR THIS PRODUCT WERE REVIEWED AND NO ANOMALIES WERE FOUND. DEVICE HISTORY RECORD (DHR) REVIEW WAS UNABLE TO BE PERFORMED AS THE LOT NUMBER OF THE DEVICE INVOLVED IN THE EVENT IS UNKNOWN. RETURNED PRODUCT WAS VISUALLY INSPECTED AND NO ANOMALIES WERE OBSERVED. FOR THESE REASONS, THE ROOT CAUSE WAS UNABLE TO BE DETERMINED. THERE IS NO EVIDENCE TO SUGGEST THAT THE CAUSE OF THE INFECTION IS IMPLANT RELATED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001032347-2018-00331-4, 0001032347-2018-00332-4, 0001032347-2018-00333-4, 0001032347-2018-00335-4, 0001032347-2018-00336-4, 0001032347-2018-00337-4.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. IT WAS REPORTED THE PATIENT DEVELOPED AN INFECTION AND "BIOFILM" AROUND THE JOINT; THEREFORE THE COMPLAINT IS CONSIDERED CONFIRMED. NO PRODUCT WAS RETURNED AND NO INSPECTIONS COULD BE PERFORMED. NO X-RAYS, SCANS, PICTURES, OR PHYSICIANS REPORTS WERE PROVIDED. NO DETAILS WERE PROVIDED AS TO THE CAUSE OF THE INFECTION. ALL TMJPMS ARE DISTRIBUTED STERILE. DEVICE HISTORY RECORD (DHR) AND STERILE CERTS OF THE TMJ-PM WERE REVIEWED AND NO DISCREPANCIES WERE FOUND. DEVICE HISTORY RECORD (DHR) REVIEW WAS UNABLE TO BE PERFORMED FOR THE SCREWS AS THE LOT NUMBER OF THE DEVICES INVOLVED IN THE EVENT IS UNKNOWN. THERE ARE NO INDICATIONS OF MANUFACTURING DEFECTS. FOR THESE REASONS, THE MOST LIKELY UNDERLYING CAUSE OF THIS COMPLAINT COULD NOT BE DETERMINED. THERE IS NO EVIDENCE TO SUGGEST THAT THE CAUSE OF THE INFECTION IS IMPLANT RELATED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001032347-2018-00331-2, 0001032347-2018-00332-2, 0001032347-2018-00333-2, 0001032347-2018-00335-2, 0001032347-2018-00336-2, AND 0001032347-2018-00337-2.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001032347-2018-00331-3, 0001032347-2018-00332-3, 0001032347-2018-00333-3, 0001032347-2018-00335-3, 0001032347-2018-00336-3, AND 0001032347-2018-00337-3.

Description of Event or Problem · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.

Description of Event or Problem · 0

THIS REPORT IS BEING SUBMITTED TO DOCUMENT THE COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 0

IT WAS REPORTED THE REMOVAL DID TAKE PLACE. THE SALES REPRESENTATIVE STATED THERE WERE NO ISSUES DURING THE REVISION TO HIS KNOWLEDGE. THE SURGEON PLANS TO SEND THE JOINT BACK TO THE MANUFACTURER ONCE THE HOSPITAL COMPLETES THEIR PROLONGED CULTURE IN THEIR LAB. THEY ESTIMATED IT WOULD BE READY TO SHIP BACK IN THREE (3) WEEKS.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET COMPLAINT (B)(4). CONCOMITANT MEDICAL PRODUCTS: BIOMET MICROFIXATION CUSTOM MADE DEVICE GEORGE PM-TMJ & MODEL CATALOG #: TMJPM-1887 LOT #: 798940A, BIOMET MICROFIXATION "2.4MM" SYSTEM HT CROSS-DRIVE SCREW CATALOG #: 91-2710 LOT #: NI, BIOMET MICROFIXATION "2.4MM" SYSTEM HT CROSS-DRIVE SCREW CATALOG #: 91-2712 LOT #: NI, BIOMET MICROFIXATION "2.4MM" SYSTEM HT CROSS-DRIVE SCREW CATALOG #: 91-2718 LOT #: NI, BIOMET MICROFIXATION TMJ SYSTEM CROSS DRIVE FOSSA SCREW CATALOG #: 99-6577 LOT #: NI, BIOMET MICROFIXATION TMJ SYSTEM CROSS DRIVE FOSSA SCREW CATALOG #: 99-6579 LOT #: NI, THERAPY DATE: (B)(6) 2018. CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION, THE DEVICES REMAIN IMPLANTED. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001032347-2018-00331, 0001032347-2018-00332, 0001032347-2018-00333, 0001032347-2018-00335, 0001032347-2018-00336, AND 0001032347-2018-00337.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001032347-2018-00331-1, 0001032347-2018-00332-1, 0001032347-2018-00333-1, 0001032347-2018-00335-1, 0001032347-2018-00336-1, AND 0001032347-2018-00337-1.

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT HAD A BIOPSY OF THE TMJ. THE SURGEON REPORTED "THE CULTURE RESULTS WERE (B)(6) OTHER THAN A FEW PMN¿S (POLYMORPHONUCLEAR LEUKOCYTES) FROM ONE OF THE THREE SITES BIOPSIED. THE SURGICAL PATH WAS ALSO (B)(6). THIS WAS DONE TO RULE OUT PJI (PERIPROSTHETIC JOINT INFECTION) AND BIOFILM DUE TO A THREE MONTH HISTORY OF INTERMITTENT LEFT FACIAL SWELLING THAT RESOLVES WITH ANTIBIOTICS." THE ONSET OF SYMPTOMS STARTED APPROXIMATELY THREE MONTHS AFTER IMPLANTATION. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.

Description of Event or Problem · 1

IT WAS ADDITIONALLY REPORTED THE SURGEON REQUESTED INFORMATION REGARDING NEW IMPLANTS FOR THE PATIENT. HE STATED THE PATIENT HAS AN INFECTION AND "BIOFILM" AND THAT HE PLANS TO REMOVE THE IMPLANTS AND PLACE SPACERS, TREAT THE PATIENT WITH ANTIBIOTICS FOR SIX WEEKS, WAIT AN ADDITIONAL TWO WEEKS TO ENSURE THE INFECTION IS CLEARED, THEN IMPLANT NEW PROSTHESIS. THE REMOVAL IS NOT YET SCHEDULED AND WILL NOT BE SCHEDULED UNTIL A TIMELINE IS ESTABLISHED FOR THE APPROVAL AND MANUFACTURE OF THE NEW PROSTHESIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
404777 "2.4MM" SYSTEM HIGH TORQUE (HT) CROSS-DRIVE SCREW BONE SCREW HWC BIOMET MICROFIXATION N/A UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 53 YR Hospitalization| R