PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS
Report
- Report Number
- 3004209178-2018-75923
- Event Type
- Injury
- Date Received
- June 1, 2018
- Date of Event
- May 11, 2018
- Report Date
- July 10, 2018
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 00643169513914
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
UNIT PASSED THE DELIVERY ACCURACY TEST. ALL OPERATING CURRENTS ARE WITHIN SPECIFICATION. NO UNEXPECTED LOW BATTERY OR OFF NO POWER ALARMS NOTED. UNIT PASSED FUNCTIONAL TEST INCLUDING DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, AND EXCESSIVE NO DELIVERY ALARM TEST. UNIT FUNCTIONED PROPERLY. NO BOLUS DELIVERY ANOMALY NOTED DURING TESTING.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WENT HOSPITAL DUE TO HIGH BLOOD GLUCOSE OF 350 MG/DL ON (B)(6) 2018. CUSTOMER¿S CURRENT BLOOD GLUCOSE WAS 201MG/DL. CUSTOMER¿S BLOOD GLUCOSE LEVEL AT THE TIME OF HOSPITALIZATION WAS 227 MG/DL. CUSTOMER HAD SYMPTOMS SUCH AS NOURISH. DRIVE SUPPORT CAP WAS NORMAL. THERE WAS NO LEAK AND AIR BUBBLE. CUSTOMER WAS ALLEGING FOR UNDER DELIVERING BECAUSE BLOOD GLUCOSE WAS NOT RESPONDING TO THE BOLUS WHICH HE GAVE. CUSTOMER EXPERIENCED BLOOD GLUCOSE LEVEL OF 240, 190, 287, MG/DL. THE CUSTOMER WAS WEARING THE INSULIN PUMP DURING THE HOSPITALIZATION. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 402795 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAS | A5723LNASJ | 00643169513914 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR | Other |