ALERE DETERMINE HIV 1/2 AG/AB COMBO
Report
- Report Number
- 1221359-2018-00106
- Event Type
- Malfunction
- Date Received
- May 30, 2018
- Date of Event
- August 17, 2016
- Report Date
- May 30, 2018
- Manufacturer
- ALERE SCARBOROUGH INC.
- Product Code
- MZF
- UDI-DI
- 07290015003735
- PMA / PMN Number
- BP120037-0
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
TESTING WAS PERFORMED ON RETAINED KIT LOT 160406 WITH THE FOLLOWING INTERNAL WHOLE BLOOD AND SERUM/PLASMA CONTROL SAMPLES: (B)(6) . THE KIT PERFORMED AS EXPECTED. ADDITIONALLY, QC DATA AND BATCH RECORDS FOR LOT 160406 WERE REVIEWED. RELEASE TESTING WAS VALID AND THE LOT WAS FOUND TO BE PERFORMING AS EXPECTED. A REVIEW OF THE COMPLAINTS REPORTED (B)(6) RELATED TO LOT NUMBER 160406 SHOWED THAT THE COMPLAINT RATE IS (B)(4). THE EVIDENCE AVAILABLE DOES NOT INDICATE THAT THE PRODUCT IS PERFORMING OUTSIDE LABEL CLAIMS. IN HOUSE TESTING WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. THE RESULTS OBTAINED MAY POSSIBLY BE RELATED TO THE PATIENT SAMPLE. THE SAMPLE MAY HAVE CONTAINED SPECIFIC SUBSTANCES WHICH MAY HAVE AFFECTED THE RESULTS. THE AVAILABLE EVIDENCE SUGGESTS THAT THIS DEVICE LOT IS PERFORMING WITHIN LABELED CLAIMS.
CUSTOMER REPORTED A CONCOMITANT MEDICAL PRODUCTS: RESULT ON AN UNKNOWN SAMPLE TYPE TESTED WITH THE ALERE DETERMINE HIV 1/2 AG/AB COMBO. (B)(6) . PATIENT GENDER, PREGNANCY STATUS, TREATMENT AND PATIENT OUTCOMES WERE UNKNOWN. AT THE TIME OF THE EVENT THERE WAS NO INDICATION OF A PRODUCT MALFUNCTION BASED ON THE INFORMATION PROVIDED AND ASSESSED BY ALERE (B)(4), INC., HOWEVER THIS MDR IS A RETROSPECTIVE FILING IN RESPONSE TO AN OBSERVATION FROM AN FDA INSPECTION CONDUCTED (B)(6) TO (B)(6) 2018 AT ALERE SCARBOROUGH, INC. ((B)(6)).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 396636 | ALERE DETERMINE HIV 1/2 AG/AB COMBO | IVD FOR HIV AG/AB | MZF | ALERE SCARBOROUGH INC. | 160406 | 07290015003735 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |