FDA Adverse Event Malfunction Summary report: N

NEPHROSCOPE 20° 24FR. WL 224MM RIGID WITH LATERAL OCULAR LENS WORKING CHANNEL

MDR report key: 7550175 · Received May 29, 2018

Report

Report Number
9611102-2018-00010
Event Type
Malfunction
Date Received
May 29, 2018
Date of Event
April 17, 2018
Report Date
May 29, 2018
Manufacturer
RICHARD WOLF GMBH
Product Code
FGA
UDI-DI
04055207023327
PMA / PMN Number
K994223
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
YM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). TO OBTAIN OPTIMUM LIGHT TRANSMISSION, THE DIAMETERS OF THE FIBRE BUNDLE OF THE ENDOSCOPE AND OF THE FIBRE LIGHT CABLE MUST BE THE SAME. POSSIBLE CAUSES WHEN USING A FIBRE LIGHT CABLE THAT IS TOO SMALL: THE LIGHT OUTPUT IS REDUCED. CAUTION! INCREASED DANGER POTENTIAL IF IMAGE IS BLURRED! PATIENT MAY BE INJURED. STOP OPERATION FOR SAFETY REASONS IF IMAGE IS BLURRED. CHECK IMAGE QUALITY OF ENDOSCOPE BEFORE USE. CAUTION! DO NOT WORK OUTSIDE THE SCOPE'S FIELD OF VIEW! INADVERTENT TISSUE DAMAGE AS WELL AS DAMAGE TO THE DISTAL END OF THE ENDOSCOPE AND INSTRUMENT PARTS CAN OCCUR. CHECK IMAGE QUALITY AND LIGHT OUTPUT IN CONJUNCTION WITH THE SYSTEM COMPONENTS. AN INVESTIGATION WAS UNABLE TO BE PERFORMED AS THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER. MULTIPLE ATTEMPTS WERE MADE TO CONTACT THE USER FACILITY, BUT NO RESPONSE WAS PROVIDED. RICHARD WOLF GMBH (RWGMBH) CONSIDERS THIS MATTER CLOSED. IN THE EVENT RWGMBH RECEIVES ADDITIONAL INFORMATION A FOLLOW UP REPORT WILL BE SUBMITTED TO FDA.

Description of Event or Problem · 0

FOLLOW UP REPORT TO PROVIDE ADDITIONAL INFORMATION TO FDA.

Additional Manufacturer Narrative · 1

INVESTIGATION / EVALUATION CURRENTLY IN PROCESS. RW (B)(4) CONSIDERS THIS REPORT OPEN, FOLLOW-UP REPORT TO BE SUBMITTED ONCE INVESTIGATION IS COMPLETE. RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) IS SUBMITTING THIS REPORT ON BEHALF OF RICHARD WOLF (B)(4). RW (B)(4) IS MANUFACTURER.

Description of Event or Problem · 1

AT THE FIRST TEN MINUTES OF THE OPERATION THE NEPHROSCOPE 8965.411 WAS EXCELLENT AND THEN WE STARTED THE SECOND OPERATION AND VISION BEGAN TO DISAPPEAR GRADUALLY THE SAME DAY. UNFORTUNATELY THE OPERATION WAS NOT COMPLETED. MANUFACTURED DATE: 08 MARCH 2018. LOT SIZE: (B)(4) UNITS. PURCHASE DATE: (B)(6) 2018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
393390 NEPHROSCOPE 20° 24FR. WL 224MM RIGID WITH LATERAL OCULAR LENS WORKING CHANNEL NEPHROSCOPE 20° 24FR WL224MM FGA RICHARD WOLF GMBH 8965.411 04055207023327

Patients

Seq Age Sex Outcome Treatment
1 21 YR Other