FDA Adverse Event Malfunction Summary report: N

SERVO-U

MDR report key: 7549283 · Received May 29, 2018

Doing adverse event research by hand?

The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.

Free to try · every answer cites its records

Report

Report Number
8010042-2018-00267
Event Type
Malfunction
Date Received
May 29, 2018
Report Date
September 21, 2018
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
CBK
PMA / PMN Number
K151814
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).

Additional Manufacturer Narrative · 0

THE VENTILATOR WAS INVESTIGATED ON SITE AND THE O2 GAS AND THE AIR GAS MODULE WERE REPLACED AND RETURNED FOR INVESTIGATION. THE O2 GAS AND AIR GAS MODULE REGULATES THE INSPIRATORY OXYGEN AND AIR GAS FLOW TO THE PATIENT. THE REPORTED ALARMS FOR HIGH O2-CONCENTRATION WERE NOT REPRODUCED DURING TEST OF THE RETURNED O2 GAS AND AIR GAS MODULE IN A REFERENCE VENTILATOR. NO ALARMS WERE GENERATED DURING VENTILATION, NOR ANY DEVIATION DURING TESTS IN THE PRODUCTION TEST SYSTEM. THE DEVICE LOGS HAVE NOT BEEN RECEIVED. THE REPORTED PROBLEM COULD NOT BE REPRODUCED, THEREFORE THE CAUSE COULD NOT BE DETERMINED IN THIS INVESTIGATION. GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC. 45 BARBOUR POND DRIVE. WAYNE, NJ 07470. CONTACT PERSON: (B)(6).

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE THE VENTILATOR WAS CONNECTED TO A PATIENT, IT GENERATED ALARMS FOR HIGH O2 CONCENTRATION. THERE WAS NO PATIENT HARM. (B)(4).

Description of Event or Problem · 0

IMPORTER REF. #: (B)(4). MANUFACTURER REF. #: (B)(4).