FDA Adverse Event Injury Summary report: N

AMISTEM-H PROXIMAL COATING STD STEM SIZE 2

MDR report key: 7549087 · Received May 29, 2018

Report

Report Number
3005180920-2018-00379
Event Type
Injury
Date Received
May 29, 2018
Date of Event
May 1, 2018
Report Date
May 29, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804373
PMA / PMN Number
K161635
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CLINICAL EVALUATION PERFORMED ON 14 MAY 2018 BY MEDICAL AFFAIRS MANAGER: FEMORAL STEM LOOSENING 8 MONTHS AFTER PRIMARY CEMENTLESS TOTAL HIP ARTHROPLASTY IN A (B)(6) YEAR OLD MAN. RADIOGRAPHIC IMAGES PROVIDED SHOW THE PRESENCE OF RADIOLUCENT LINES IN GRUEN ZONES 1, 2, 6 AND 7 AND SIGNS OF STRESS SHIELDING. ASEPTIC LOOSENING IS A POSSIBLE, LITERATURE DESCRIBED ADVERSE EVENT AFTER CEMENTLESS THA AND CAUSES ARE OFTEN UNKNOWN. THE REASON OF THIS FAILURE CANNOT BE DETERMINED BATCH REVIEW PERFORMED ON 28 MAY 2018; LOT 161299: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 08 APRIL 2016. EXPIRATION DATE: 2021-03-29 NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. VISUAL INSPECTION PERFORMED ON 28 MAY 2018 BY R&D PRODUCT MANAGER: THE COATING OF THE STEM APPEARED ALMOST TOTALLY PRESENT IN THE STEM: THE PRESENCE OF THE COATING CAN INDICATE AN INCORRECT OSSEOINTEGRATION WITH THE FEMUR, ALTHOUGH SIGNS OF REGROWTH BONE WERE FOUND IN THE DISTAL PART OF THE COATING. MOREOVER, SOME SMALL SIGNS OF REMOVAL WERE FOUND IN THE TAPER AREA. FROM THE RECEIVED PIECE IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF THE EVENT.

Description of Event or Problem · 1

ABOUT 8 MONTHS AFTER PRIMARY SURGERY THE SURGEON REVISED THE PATIENT HIP DUE TO STEM LOOSENING. THE SURGEON SWAPPED SUCCESSFULLY THE STEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
392081 AMISTEM-H PROXIMAL COATING STD STEM SIZE 2 CEMENTLESS HIP STEM LZO MEDACTA INTERNATIONAL SA 161299 07630030804373

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention