FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 7545756 · Received May 25, 2018

Report

Report Number
3008642652-2018-04743
Event Type
Malfunction
Date Received
May 25, 2018
Date of Event
May 10, 2018
Report Date
May 24, 2018
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
UDI-DI
00855778005012
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM THAT THE (THERAPY ELECTRODES WERE NON-FUNCTIONAL) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT FAILED A THERAPY ELECTRODE RECOGNITION TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO AN OPEN PULSE WIRE IN THE CABLE CONNECTING ECG "A" AND ECB "B". THE ROOT CAUSE FOR THE OPEN WIRE WAS EXCESSIVE FORCE. BELT IS DESIGNED TO MEET THE MECHANICAL REQUIREMENTS OF IEC (B)(4). NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT.

Description of Event or Problem · 1

A US DISTRIBUTOR RETURNED A PATIENT'S ELECTRODE BELT AND REPORTED THAT THE THERAPY ELECTRODES WERE NON-FUNCTIONAL.