FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 7545756
·
Received May 25, 2018
Report
- Report Number
- 3008642652-2018-04743
- Event Type
- Malfunction
- Date Received
- May 25, 2018
- Date of Event
- May 10, 2018
- Report Date
- May 24, 2018
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- UDI-DI
- 00855778005012
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM THAT THE (THERAPY ELECTRODES WERE NON-FUNCTIONAL) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT FAILED A THERAPY ELECTRODE RECOGNITION TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO AN OPEN PULSE WIRE IN THE CABLE CONNECTING ECG "A" AND ECB "B". THE ROOT CAUSE FOR THE OPEN WIRE WAS EXCESSIVE FORCE. BELT IS DESIGNED TO MEET THE MECHANICAL REQUIREMENTS OF IEC (B)(4). NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT.
Description of Event or Problem · 1
A US DISTRIBUTOR RETURNED A PATIENT'S ELECTRODE BELT AND REPORTED THAT THE THERAPY ELECTRODES WERE NON-FUNCTIONAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 388846 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 | 00855778005012 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |