FDA Adverse Event Malfunction Summary report: N

ALERE DETERMINE HIV 1/2 AG/AB COMBO

MDR report key: 7542791 · Received May 24, 2018

Report

Report Number
1221359-2018-00039
Event Type
Malfunction
Date Received
May 24, 2018
Date of Event
August 12, 2016
Report Date
May 24, 2018
Manufacturer
ALERE SCARBOROUGH INC.
Product Code
MZF
UDI-DI
07290015003735
PMA / PMN Number
BP120037-0
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IN-HOUSE TESTING WAS PERFORMED ON RETAINED KIT LOT 160406 WITH THE FOLLOWING INTERNAL WHOLE BLOOD CONTROL SAMPLES:(B)(6). ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING BATCH RECORDS FOR LOT 160406 WERE REVIEWED. THIS LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED (B)(6) RELATED TO LOT NUMBER 160406 SHOWED THAT THE COMPLAINT RATE IS (B)(4). THE EVIDENCE AVAILABLE DOES NOT INDICATE THAT THE PRODUCT IS PERFORMING OUTSIDE LABEL CLAIMS. THE EXACT ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. THE RESULTS OBTAINED MAY POSSIBLY BE RELATED TO THE PATIENT SAMPLE. THE SAMPLE MAY HAVE CONTAINED SPECIFIC SUBSTANCES WHICH MAY HAVE AFFECTED THE RESULTS. THE AVAILABLE EVIDENCE SUGGESTS THAT THIS DEVICE LOT IS PERFORMING WITHIN LABELED CLAIMS.

Description of Event or Problem · 1

A (B)(6) RESULT WAS REPORTED ON A FINGERSTICK BLOOD SAMPLE TESTED WITH THE ALERE DETERMINE HIV 1/2 AG/AB COMBO. A SECOND TEST WAS COMPLETED WITH (B)(6) AND THE RESULT WAS (B)(6). THE PATIENT WAS CONFIRMED (B)(6). THERE IS INSUFFICIENT INFORMATION TO DETERMINE IF A MALFUNCTION OCCURRED. THE PATIENT GENDER, PREGNANCY STATUS, TREATMENT AND PATIENT OUTCOME WERE UNKNOWN. AT THE TIME OF THE EVENT THERE WAS NO INDICATION OF A PRODUCT MALFUNCTION BASED ON THE INFORMATION PROVIDED AND ASSESSED BY ALERE (B)(4), INC.; HOWEVER, THIS MDR IS A RETROSPECTIVE FILING IN RESPONSE TO AN OBSERVATION FROM AN FDA INSPECTION CONDUCTED (B)(6) TO (B)(6), 2018 AT ALERE (B)(4), INC. ((B)(4)).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
384733 ALERE DETERMINE HIV 1/2 AG/AB COMBO IVD FOR HIV AG/AB MZF ALERE SCARBOROUGH INC. 160406 07290015003735

Patients

Seq Age Sex Outcome Treatment
1