FDA Adverse Event Injury Summary report: N

G7 ACETABULAR 3 HOLE SHELL

MDR report key: 7537076 · Received May 23, 2018

Report

Report Number
0001825034-2018-03547
Event Type
Injury
Date Received
May 23, 2018
Date of Event
April 25, 2018
Report Date
September 27, 2018
Manufacturer
ZIMMER BIOMET, INC.
Product Code
PBI
PMA / PMN Number
PK121874
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS NOT CONFIRMED. ACETABULAR SHELL WAS RETURNED AND EVALUATED AGAINST THE COMPLAINT. VISUAL INSPECTION FOUND SCRATCHING ON THE INNER RADIUS OF THE SHELL. THE INNER SIDE WALL OF THE SHELL WAS ALSO SCRATCHED. NO DAMAGE WAS OBSERVED TO THE LOCKING GROOVE. DEVICE MET ALL DIMENSIONAL SPECIFICATIONS. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 1

(B)(4). CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: THE 010000658 G7 ACETABULAR SHELL 3930175. THE 010000768 G7 ARCOM LINER 6109148. THE 010000913 G7 HI-WALL E1 LINER 6167725. THE 103530 TI PROFILE SCREW 2016040532. THE 103532 TI PROFILE SCREW 2015090213. THE 103533 TI PROFILE SCREW 2017011260. THE 130914 MULLER CUP 3753738. THE 00801802801 VERSYS FEMORAL HEAD 63920806. THE 0100561213 WAGNER TAPER STEM 2761649. (B)(6). MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2018-03550, 0001825034-2018-03551, 0001825034-2018-03552.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A HIP ARTHROPLASTY THE POLYETHYLENE LINER WOULD NOT SEAT WITH THE ACETABULAR SHELL. THERE WAS AN ADDED DELAY OF ONE-TWO HOURS. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
378916 G7 ACETABULAR 3 HOLE SHELL PROSTHESIS, HIP PBI ZIMMER BIOMET, INC. N/A 6122056

Patients

Seq Age Sex Outcome Treatment
1 64 YR Required Intervention