TITAN TOUCH SCRO ZERO ANG 22CM
Report
- Report Number
- 2125050-2018-00401
- Event Type
- Injury
- Date Received
- May 22, 2018
- Date of Event
- April 30, 2018
- Report Date
- May 22, 2018
- Manufacturer
- COLOPLAST A/S
- Product Code
- FHW
- UDI-DI
- 05708932539203
- PMA / PMN Number
- P000006
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE DEVICE IS NOT AVAILABLE FOR EVALUATION. WITHOUT THE BENEFIT OF ANALYZING THE EXPLANT, QUALITY CANNOT CONFIRM ANY OBSERVATIONS AND CANNOT COMMENT ON THE CONDITION OF THE PROSTHESIS. QUALITY REVIEWED THE MANUFACTURING RECORDS FOR THIS DEVICE AND CONFIRMED THAT THERE WERE NO DISCREPANCIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT AND VERIFIED THAT THE DEVICES FROM THIS LOT MET ALL SPECIFICATIONS PRIOR TO RELEASE. A REVIEW OF THE COMPLAINT HISTORY DATABASE REVEALED NO ADDITIONAL REPORTS OR TRENDS OF INCORRECT SIZING ASSOCIATED WITH LOT NUMBER 4904043. REVIEW OF NONCONFORMANCES REVEALED NO NONCONFORMANCE WITH THIS LOT THAT WOULD HAVE CONTRIBUTED TO THE EVENT. THE MOST LIKELY ROOT CAUSE OF THE EVENT MAY BE ATTRIBUTED TO INCORRECT SIZING OF THE PATIENT DURING THE IMPLANT PROCEDURE. IF ADDITIONAL INFORMATION IS RECEIVED, QUALITY WILL RE-EVALUATE THIS COMPLAINT IN ACCORDANCE TO PROCEDURES. MANAGEMENT ROUTINELY REVIEWS EVENTS SUCH AS THIS AND MONITORS COMPLAINT LEVELS. ADDITIONALLY, EVENTS OF THIS TYPE ARE CAPTURED IN THE PRODUCT RISK DOCUMENTATION. BASED ON THIS INFORMATION, NO FURTHER CORRECTIVE ACTION IS REQUIRED AT THIS TIME.
ACCORDING TO THE AVAILABLE INFORMATION, THE PHYSICIAN SHORTENED THE TUBING LENGTH AND DID A WASHOUT. REPOSITIONED THE SAME PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 374774 | TITAN TOUCH SCRO ZERO ANG 22CM | INFLATABLE PENILE PROSTHESIS | FHW | COLOPLAST A/S | ES29222400 | 4904043 | 05708932539203 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Other |