FDA Adverse Event Malfunction Summary report: N

BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER

MDR report key: 7534558 · Received May 22, 2018

Report

Report Number
9610048-2018-00060
Event Type
Malfunction
Date Received
May 22, 2018
Date of Event
May 5, 2018
Report Date
May 31, 2018
Manufacturer
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
Product Code
FOZ
UDI-DI
00382903818235
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: SAMPLES/ PHOTOS: SAMPLES HAVE NOT RECEIVED FOR ANALYSIS OF THE INCIDENT IN QUESTION. DHR REVIEW: THE FOLLOWING ASSEMBLY BATCH USED IN THE FINAL CLAIMED PRODUCT LIST 7244583 WAS ANALYZED: THE FINAL ASSEMBLY LOT: 7241767 ASSEMBLED IN THE PERIOD FROM 09/01/2017 TO 09/13/2017. THE BATCH WAS ANALYZED FOR "NEEDLE RETRACTION" AND ¿PART ACTIVATION¿ TESTS, AND IT WAS NOT EVIDENCED RECORD OF FAILURE OF ACTIVATION OF THE PART DURING THE ANALYSIS OF THESE BATCHES. QN/ NCMR REVIEW: NO RECORDS OF QUALITY NOTIFICATION OR REPORTS OF NON-CONFORMITY THAT COULD LEAD TO INCIDENTS IN TO THE LOT INVOLVED IN THIS COMPLAINT WERE EVIDENCED. THE CORRECTIVE MAINTENANCE HISTORY IN THE MANUFACTURING PERIOD OF THE ASSEMBLY LOT INVOLVED IN THIS COMPLAINT WAS EVALUATED AND NO RECORDS WERE RECORDED THAT COULD CLEARLY INDICATE A FAILURE OF THE ACTIVATION OF THE PART. INVESTIGATION CONCLUSION: NOT CONFIRMED: BD WAS UNABLE TO CONFIRM THE CLAIMED DEFECT. SINCE SAMPLES CONTAINING THE DEFECT CLAIMED ARE NOT AVAILABLE FOR REVIEW, THIS COMPLAINT WAS NOT CONFIRMED. IN ADDITION, THERE ARE NO NON-CONFORMITIES OR CORRECTIVE MAINTENANCE IN THE LOTS INVOLVED IN THIS COMPLAINT, WHICH COULD CONFIRM THE DEFECT CLAIMED WITHOUT PRESENCE OF SAMPLES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SAFETY MECHANISM OF A BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER WAS NOT WORKING PROPERLY. THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
378208 BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER INTRAVASCULAR CATHETER FOZ BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. 7244592 00382903818235

Patients

Seq Age Sex Outcome Treatment
1 Other