GLASSVAN
Report
- Report Number
- 1058382-2018-00004
- Event Type
- Malfunction
- Date Received
- May 22, 2018
- Date of Event
- May 14, 2018
- Report Date
- May 21, 2018
- Manufacturer
- M/S NIRAJ INDUSTRIES PVT LTD.
- Product Code
- GES
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- SC, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE MANUFACTURER HAS INVESTIGATED THIS ISSUE ON THE CONTROL SAMPLES AND SAMPLES PROVIDED BY THE COMPLAINANT OF THE SAME BATCH NUMBER (171611AC). THE REPORTED PROBLEM COULD NOT BE ESTABLISHED ON EXAMINATION OF RETAINED SAMPLES AND RELATED RECORDS OF BATCH NUMBER 171611AC. THE DEVICE CHARACTERISTICS WHICH CAN INDUCE BREAKAGE OF BLADE HAVE BEEN OBSERVED WITHIN SPECIFICATIONS. 34 UNUSED SAMPLES PROVIDED BY THE COMPLAINANT ARE ALSO FOUND WITHIN THE SPECIFICATIONS WITH REGARD TO THE PARAMETERS WHICH CAN INDUCE BLADE BREAKAGE (SHARPNESS, HARDNESS, STIFFNESS AND FITMENT OF BLADES). SAMPLES EVALUATION REPORTS ARE ENCLOSED FOR YOUR READY REFERENCE. WE HAVE CLOSED THE COMPLAINT BASED ON THE RESULT OF THE MANUFACTURER'S INVESTIGATION. WE CONSIDER THIS REPORT CLOSED.
IT WAS REPORTED THAT THE #15 SURGICAL BLADE BROKE DURING THE REMOVAL OF IMPACTED 3RD MOLARS OF THE FEMALE PATIENT. THE BROKEN BLADE (FRAGMENT) WAS QUICKLY RECOVERED FROM THE PATIENT'S MOUTH AND NO FURTHER MEDICAL INTERVENTION OR ADDITIONAL TREATMENT WAS NEEDED DUE TO THIS ISSUE.
IT WAS REPORTED THAT THE #15 SURGICAL BLADE BROKE DURING THE REMOVAL OF IMPACTED 3RD MOLARS OF THE FEMALE PATIENT. THE BROKEN BLADE (FRAGMENT) WAS QUICKLY RECOVERED FROM THE PATIENT'S MOUTH AND NO FURTHER MEDICAL INTERVENTION OR ADDITIONAL TREATMENT WAS NEEDED DUE TO THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 375692 | GLASSVAN | SURGICAL BLADE | GES | M/S NIRAJ INDUSTRIES PVT LTD. | 171611AC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |