FDA Adverse Event Malfunction Summary report: N

BD MICRO FINE PLUS ¿ PEN INJECTOR NEEDLE

MDR report key: 7531788 · Received May 22, 2018

Report

Report Number
9616656-2018-00113
Event Type
Malfunction
Date Received
May 22, 2018
Date of Event
May 10, 2018
Report Date
June 14, 2018
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: TWO OPEN AND FOURTEEN SEALED 32G X 6MM PEN NEEDLE SAMPLES ALONG WITH FOUR PHOTOS WERE RETURNED FROM LOT. NO. 7172737, CAT. NO. 320738. VISUAL EXAMINATION OF RETURNED SAMPLES AND PHOTOS WAS CARRIED OUT AND A BROKEN NON PATIENT END OF NEEDLE WAS OBSERVED ON ONE OPENED SAMPLE, NO ISSUES WERE OBSERVED WITH THE SECOND OPEN SAMPLE OR THE FOURTEEN SEALED SAMPLES. INVESTIGATION CONCLUSION: VISUAL EXAMINATION OF RETURNED SAMPLES AND PHOTOS WAS CARRIED OUT AND A BROKEN NON PATIENT END OF NEEDLE WAS OBSERVED ON ONE OPENED SAMPLE, NO ISSUES WERE OBSERVED WITH THE SECOND OPEN SAMPLE OR THE FOURTEEN SEALED SAMPLES. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED WITH THE USE OF THE BD MICRO FINE PLUS¿ PEN INJECTOR NEEDLE THERE WAS AN ISSUE WITH A BROKEN NEEDLE. IT WAS STATED ¿WHEN INJECTED, THE PATIENT COULDN'T PRESS THE BUTTON AND THE NEEDLE WAS BROKEN WHEN REMOVED FROM THE SKIN". THERE WAS NO REPORT OF INJURY OR FURTHER MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
378146 BD MICRO FINE PLUS ¿ PEN INJECTOR NEEDLE PEN INJECTOR NEEDLE FMI BECTON DICKINSON AND CO. 7172737

Patients

Seq Age Sex Outcome Treatment
1 Other