A1A-600II
Report
- Report Number
- 8031673-2018-00461
- Event Type
- Malfunction
- Date Received
- May 19, 2018
- Date of Event
- May 3, 2018
- Report Date
- May 19, 2018
- Manufacturer
- TOSOH CORPORATION
- Product Code
- KHO
- PMA / PMN Number
- K971103
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
TOSOH BIOSCIENCE, INC. IS SUBMITTING ON BEHALF OF THE FOREIGN MANUFACTURER, TOSOH CORPORATION, PER EXEMPTION NUMBER E2017013. FIELD SERVICE ENGINEERING (FSE) VISITED THE CUSTOMER TO ADDRESS THE REPORTED EVENT. DURING TROUBLESHOOTING, FSE FOUND THE SUBSTRATE NOZZLE IN THE WRONG LOCATION AND THE PROBE ASSAY TIP WAS CRUNCHED PREVENTING A TIP FROM ATTACHING TO IT. FSE REPLACED EACH WASH PROBE AND ALSO CLEANED AND CORRECTLY PLACED THE SUBSTRATE NOZZLE. TO VERIFY THE INSTRUMENT, FSE THEN RAN THE DAILY CHECK WITHOUT ANY ISSUE. THEY ALSO RAN 5 SAMPLES OF TOTAL T3 (TT3) WITH ACCEPTABLE RESULTS. THE INSTRUMENT WAS VERIFIED AS OPERATIONAL. THERE WAS NO FURTHER ACTION REQUIRED BY FSE. A COMPLAINT HISTORY REVIEW AND SERVICE HISTORY REVIEW FOR SIMILAR COMPLAINTS WAS PERFORMED ON (B)(4) 2018 FOR A1A-600 INSTRUMENT WITH SERIAL NUMBER (B)(4). THE SEARCH WAS CONDUCTED FROM (B)(4) 2017 THROUGH AWARE DATE (B)(4)2018 AND DID NOT FIND ANY SIMILAR COMPLAINTS DURING THE SEARCHED PERIOD. THE AIA-600 OPERATOR'S MANUAL UNDER CHAPTER 8- TROUBLESHOOTING STATES THAT A DL FLAG IS CAUSED BY THE FOLLOWING: OUT OF SUBSTRATE II, OR RUNNING WITH CLEANING SOLUTION. LOW DETECTOR LAMP INTENSITY. TO RESOLVE THE ISSUE, IT INDICATES THAT THE FOLLOWING MUST BE DONE: REPLENISH SUBSTRATE II, PERFORM SUBSTRATE BACKGROUND CHECK. CALL TOSOH BIOSCIENCE. THE MOST PROBABLE CAUSE OF THE DETECTOR LOW (DL) FLAG WAS A DAMAGED WASH PROBE TIP AND THE SAMPLE NOZZLE INCORRECTLY PLACED.
A CUSTOMER REPORTED DETECTOR LOW (DL) FLAGS WITH THE A1A-600 INSTRUMENT. TECHNICAL SUPPORT ADVISED THE CUSTOMER TO MAKE UP NEW SUBSTRATE BUT THEY CONTINUED TO HAVE THE DL FLAGS. THE ANALYZER WAS DOWN. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ADDRESS THE REPORTED EVENT, WHICH RESULTED IN DELAYED REPORTING OF BETA HUMAN CHORIONIC GONADOTROPIN (BHCG) AND PARATHYROID HORMONE (IPTH) PATIENT RESULTS. THERE WAS NO INDICATION OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DELAY IN REPORTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 370849 | A1A-600II | A1A-600II | KHO | TOSOH CORPORATION | A1A-600II |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |