FDA Adverse Event Other Summary report: N

BITE PROOF BITE BLOCK

MDR report key: 752654 · Received May 11, 2005

Report

Report Number
MW1035453
Event Type
Other
Date Received
May 11, 2005
Date of Event
May 4, 2005
Report Date
May 11, 2005
Manufacturer
B& B MEDICAL TECHNOLOGIES
Product Code
JXL
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

PATIENT INTUBATED FOR PNEUMONIA. A BITE BLOCK WAS PLACED COINCIDENT WITH ETT PLACEMENT. TWO WEEKS AFTERWARD A PERSISTENT LEAK WAS NOTED WHICH COULD NOT BE RESOLVED WITH REINFLATION OF THE ETT CUFF AND ONLY 3 CC OF AIR COULD BE INFLATED AT ANY TIME. THE PATIENT EXPERIENCED OXYGEN DESATURATION INTERMITTENTLY BUT NO OTHER SEQUELAE. THE PATIENT WAS SCHEDULED FOR TRACHEOSTOMY BECAUSE OF PROLONGED INTUBATION. IT WAS ASSUMED THAT THE ETT CUFF WAS MALFUNCTIONING. AT TRACHEOSTOMY, THE ETT WAS REMOVED REVEALING OBSTRUCTION BY THE BITE BLOCK OF THE SOFT PLASTIC TUBING BETWEEN THE PILOT BALLOON AND THE ETT CUFF. THIS OCCURRED WITH USUAL PLACEMENT OF THIS COMBINATION OF THE ETT AND BITE BLOCK. THIS CAUSED DIFFICULTY FULLY INFLATING AND DEFLATING THE ETT CUFF WITH RESULTANT INCOMPLETE SEAL OF THE CUFF WITH THE TRACHEA AND AN AIR LEAK AS WELL AS AN INCOMPLETELY DEFLATED ETT CUFF UPON EXTUBATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BITE PROOF BITE BLOCK BITE BLOCK JXL B& B MEDICAL TECHNOLOGIES * 000105

Patients

Seq Age Sex Outcome Treatment
1 54 YR Other NOT RELEVENT