FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® CPT GLASS MOLECULAR DIAGNOSTICS TUBE

MDR report key: 7526059 · Received May 18, 2018

Report

Report Number
1917413-2018-00500
Event Type
Malfunction
Date Received
May 18, 2018
Date of Event
April 30, 2018
Report Date
June 5, 2018
Manufacturer
BECTON, DICKINSON & CO.
Product Code
JKA
UDI-DI
50382903627535
PMA / PMN Number
K891407
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES, BUT PHOTOS WERE PROVIDED BY THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR POOR BARRIER WITH THE INCIDENT LOT WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES OF THIS ISSUE ARE IDENTIFIED. INVESTIGATION CONCLUSION: BASED ON EVALUATION OF THE CUSTOMER PHOTOS, THE CUSTOMER¿S INDICATED FAILURE MODE FOR POOR BARRIER WITH THE INCIDENT LOT WAS OBSERVED. FURTHER INVESTIGATION HAS BEEN INITIATED THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES ARE IDENTIFIED. ROOT CAUSE DESCRIPTION: A CAPA HAS BEEN INITIATED TO DOCUMENT FURTHER INVESTIGATION AND ROOT CAUSE ANALYSIS RELATING TO THIS ISSUE. THE INVESTIGATION IS CURRENTLY ON-GOING AND WILL BE UPDATED AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED. CAPA (B)(4). RATIONALE: BASED ON AN ASSESSMENT OF SEVERITY AND FREQUENCY, IT WAS DETERMINED THAT A CAPA IS REQUIRED AT THIS TIME IN ORDER TO DETERMINE THE ROOT CAUSE ASSOCIATED WITH THIS ISSUE AND IMPLEMENT CORRECTIVE ACTIONS. THE INVESTIGATION IS CURRENTLY ON-GOING AND FURTHER IMPROVEMENTS WILL BE MADE AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 7251697, MEDICAL DEVICE EXPIRATION DATE: 2018-09-30, DEVICE MANUFACTURE DATE: 2017-09-08, MEDICAL DEVICE LOT #: 8004741, MEDICAL DEVICE EXPIRATION DATE: 2019-01-31, DEVICE MANUFACTURE DATE: 2018-01-04.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE A BD VACUTAINER® CPT GLASS MOLECULAR DIAGNOSTICS TUBE MALFUNCTIONED AS" ONE OF OUR SITES COLLECTED THE BLOOD AND SPUN THE TUBES AT 4000RPM BEFORE SHIPPING IT TO US AMBIENT. THE TUBE WERE RECEIVED BY OUR CENTRAL LAB (<24 HRS) AND THE GEL GOT DISASSOCIATED DURING THE TRANSIT." THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION NEEDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
370590 BD VACUTAINER® CPT GLASS MOLECULAR DIAGNOSTICS TUBE BLOOD COLLECTION TUBE JKA BECTON, DICKINSON & CO. SEE H.10 50382903627535

Patients

Seq Age Sex Outcome Treatment
1 Other