VISIONAIRE
Report
- Report Number
- 3004972304-2018-00016
- Event Type
- Malfunction
- Date Received
- May 16, 2018
- Date of Event
- April 14, 2018
- Report Date
- August 8, 2019
- Manufacturer
- CAIRE INC.
- Product Code
- CAW
- PMA / PMN Number
- K872534
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
THE VISIONAIRE STATIONARY CONCENTRATOR IN QUESTION WAS NOT EVALUATED DIRECTLY. INVESTIGATION OF THE ADVERSE EVENT WAS CONDUCTED USING A HISTORICAL ANALYSIS. REFER TO CAPA (B)(4). BASED ON THE INCIDENT DESCRIPTION IN THE ADVERSE EVENT FORM, THE VISIONAIRE UNIT'S FINAL PART NUMBER, AS098-1, AND THE VISIONAIRE UNIT'S SERIAL NUMBER, (B)(4), IT WAS DETERMINED THAT THE ADVERSE EVENT FALLS WITHIN THE PURVIEW OF CAPA (B)(4). THE CAPA ADDRESSES INCIDENTS IN WHICH FLUID ENTERING THE CONCENTRATOR UNIT CAUSED THE UNIT'S ON/OFF SWITCH TO SHORT. THE RELATED INVESTIGATION POINTED TO IMPROPER CLEANING CAUSING FLUID TO ENTER THE UNIT AND COME IN CONTACT WITH THE ON/OFF SWITCH. THE CAPA DETERMINED THAT THE RISK FACTOR IS LOW AND SEVERAL ACTIONS WERE TAKEN TO MITIGATE THE FAILURE.
THE UNIT HAS NOT BEEN RETURNED FOR EVALUATION. IF ANY NEW INFORMATION IS DISCOVERED, A FOLLOW-UP REPORT WILL BE SUBMITTED.
UNIT WAS RUNNING ON PATIENT. THEN STARTED ALARMING. THE UNIT SMELLED LIKE IT WAS BURNING. THE NURSE SWAPPED IT OUT FOR A DIFFERENT UNIT. THE UNIT WAS SITTING IN A HALL UNPLUGGED AND OBSERVED TO BE SMOKING. THE NURSE WHEELED THE UNIT OUTSIDE DURING WHICH A SMALL FLAME APPEARED AT THE SWITCH. THE FLAME WAS EXTINGUISHED WITH EXTINGUISHER. UNIT WAS RAINED ON AFTER EVENT WHILE OUTSIDE. THE WHOLE SCENARIO TOOK ABOUT 5 MINS. THE TECH IS UNSURE IF THE UNIT MALFUNCTIONED OR IF EXCESSIVE CLEANING SOLUTION MAY HAVE CAUSED THE SWITCH TO SHORT OUT
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 363244 | VISIONAIRE | OXYGEN CONCENTRATOR, STATIONARY | CAW | CAIRE INC. | AS098-1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |