FDA Adverse Event Malfunction Summary report: N

SERVO-S

MDR report key: 7518773 · Received May 16, 2018

Report

Report Number
8010042-2018-00236
Event Type
Malfunction
Date Received
May 16, 2018
Date of Event
April 30, 2018
Report Date
September 5, 2018
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
CBK
PMA / PMN Number
K123149
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE VENTILATOR WAS INVESTIGATED ON SITE AND AN O2 GAS MODULE WAS REPLACED AND RETURNED FOR INVESTIGATION. THE REPORTED FLOW TRANSDUCER TEST IN PRE-USE CHECK WAS REPRODUCED DURING TEST OF THE RETURNED GOODS IN A REFERENCE VENTILATOR. THERE WAS FOUND THAT THE OXYGEN GAS MODULE HAD A MALFUNCTION IN THE ELECTRONICS THAT ARE RESPONSIBLE FOR THE REGULATION OF OXYGEN GAS FLOW. IF THIS FAULT HAPPENS DURING VENTILATION, THE PATIENT MAY RECEIVE AN AMOUNT OF OXYGEN IN THE GAS MIXTURE THAT DEVIATE FROM THE EXPECTED. FLOW AND PRESSURE MAY ALSO DEVIATE FROM THE EXPECTED. THE FAILURE WAS CONFIRMED IN RECEIVED DEVICE LOGS. WE COULD TRACE BACK THE REPORTED PROBLEM TO (B)(6) THEREFORE THE DATE OF EVENT WAS DETERMINED AS (B)(6) 2018. GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).

Description of Event or Problem · 0

(B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE VENTILATOR FAILED THE FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER REF. #: (B)(4).