TITAN TOUCH PUMP
Report
- Report Number
- 2125050-2018-00369
- Event Type
- Injury
- Date Received
- May 15, 2018
- Date of Event
- April 24, 2018
- Report Date
- June 18, 2018
- Manufacturer
- COLOPLAST A/S
- Product Code
- FHW
- UDI-DI
- 05708932533966
- PMA / PMN Number
- P000006
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
THIS FOLLOW-UP MDR IS CREATED TO DOCUMENT THE EVALUATION OF THE RETURNED DEVICE. A TITAN TOUCH PUMP WAS RECEIVED FOR EVALUATION. EXAMINATION AND TESTING OF THE RETURNED COMPONENT REVEALED NO FUNCTIONAL ABNORMALITIES WITH THE PUMP. THE INFORMATION RECEIVED INDICATED A HARD PUMP, BUT BECAUSE NO FUNCTIONAL ABNORMALITIES WERE NOTED WITH THE RETURNED COMPONENTS, QUALITY CANNOT CONFIRM THE COMPLAINT AS REPORTED. QUALITY REVIEWED THE MANUFACTURING RECORDS FOR THIS DEVICE AND CONFIRMED THAT THERE WERE NO DISCREPANCIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT AND VERIFIED THAT THE DEVICES FROM THIS LOT MET ALL SPECIFICATIONS PRIOR TO RELEASE. A REVIEW OF THE COMPLAINT HISTORY DATABASE REVEALED NO ADDITIONAL COMPLAINTS FOR LOT 5813255. REVIEW OF NONCONFORMING REPORTS REVEALED NO NONCONFORMANCE'S FOR THIS LOT. NO CAPAS ARE ASSOCIATED WITH THIS LOT. MANAGEMENT ROUTINELY REVIEWS EVENTS SUCH AS THIS AND MONITORS COMPLAINT LEVELS. ADDITIONALLY, EVENTS OF THIS TYPE ARE CAPTURED IN THE PRODUCT RISK DOCUMENTATION. BASED ON THIS INFORMATION, NO FURTHER CORRECTIVE ACTION IS REQUIRED AT THIS TIME.
THE DEVICE HAS BEEN RECEIVED AT COLOPLAST; HOWEVER THE EVALUATION IS NOT YET COMPLETE. WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCURRENCE. ONCE OUR EVALUATION IS COMPLETE, A FOLLOW-UP REPORT WILL BE SUBMITTED. QUALITY REVIEWED THE MANUFACTURING RECORDS FOR THIS DEVICE AND CONFIRMED THAT THERE WERE NO DISCREPANCIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT AND VERIFIED THAT THE DEVICES FROM THIS LOT MET ALL SPECIFICATIONS PRIOR TO RELEASE.
ACCORDING TO THE AVAILABLE INFORMATION, PUMP NOT CYCLING. DESCRIBED AS BULB WILL NOT DEPRESS, PUMP GETS STUCK. DOCTOR INDICATED THAT SHE HAD TO REALLY WORK THE PUMP TO GET IT GOING. THE DEVICE WAS EXPLANTED ON (B)(6) 2018.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 356425 | TITAN TOUCH PUMP | INFLATABLE PENILE PROSTHESIS | FHW | COLOPLAST A/S | 5177502400 | 5813255 | 05708932533966 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Other |