HEARTSINE SAMARITAN 300P AND PAD-PAK
Report
- Report Number
- 3004123209-2018-00331
- Event Type
- Malfunction
- Date Received
- May 15, 2018
- Date of Event
- April 25, 2018
- Report Date
- June 20, 2018
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- MKJ
- PMA / PMN Number
- K014067
- Removal / Correction Number
- Z-0124-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). H3 OTHER TEXT : DEVICE NOT RETURNED YET.
THE DEVICE HISTORY RECORDS FOR THE SAM 300P DEVICE WAS REVIEWED AND THIS CONFIRMED THAT ALL MANUFACTURING AND QUALITY CHECKS AND TESTS HAD BEEN SUCCESSFULLY COMPLETED. NO REWORK WAS CONDUCTED. THE DEVICE WAS RELEASED WITH AN APPLICABLE CONCESSION, REFERENCE (B)(4). THIS CONCESSION ALLOWED FOR THE ADDITION OF A FOAM INSULATION STRIP TO PREVENT SHORTING BETWEEN CONTACTS ON THE SPEAKER. THIS CONCESSION WAS LIMITED TO THE 30TH AUGUST 2009. THE SAM 300P PASSED ¿OUT QAT FROM HEARTSINE TECHNOLOGIES ON THE 5TH AUGUST 2009. THE REPORTED PROBLEM, OF THE DEVICE SWITCHES ON AUTOMATICALLY, WAS ATTRIBUTED TO A MEMBRANE FAILURE. THE RANDOM NATURE OF THE EVENTS STORED IN THE MEMORY AND THE 10 MINUTE TIME OUTS, WOULD SUGGEST A FAILING MEMBRANE. EXPERIENCE HAS SHOWN THAT IT IS REASONABLE TO CONCLUDE THAT THESE SYMPTOMS MAY BE ATTRIBUTED TO A MEMBRANE FAILURE. THE RETURNED SAM 300P IS NOW OUTSIDE OF ITS WARRANTY PERIOD AND WILL BE SCRAPPED.
DEVICE SWITCHING ON AUTOMATICALLY. NO PATIENT INVOLVED.
DEVICE SWITCHING ON AUTOMATICALLY. NO PATIENT INVOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 355825 | HEARTSINE SAMARITAN 300P AND PAD-PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES LTD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |