FDA Adverse Event Malfunction Summary report: N

BD ULTRA-FINE INSULIN SYRINGE AND NEEDLE

MDR report key: 7512744 · Received May 14, 2018

Report

Report Number
1920898-2018-00352
Event Type
Malfunction
Date Received
May 14, 2018
Date of Event
April 24, 2018
Report Date
June 15, 2018
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 0

TWO INVESTIGATIONS WERE CARRIED OUT. FIRST INVESTIGATION SUMMARY: CUSTOMER RETURNED (10) 1/2CC, 8MM, 30G SYRINGES IN AN OPEN POLY BAG FROM LOT # 7100858. CUSTOMER STATES THAT THE SYRINGE DID NOT PULL THE MEDICINE AND THE CYLINDER WAS CRACKED. ALL RETURNED SYRINGES WERE EXAMINED AND ONE SAMPLE EXHIBITED A CRACK IN THE BARREL RANGING FROM THE 3-15 UNIT MARKINGS. THIS COULD CAUSE THE SYRINGE TO NOT DRAW PROPERLY. ALL REMAINING SYRINGES WERE TESTED AND ALL WERE ABLE TO DRAW AND EXPEL PROPERLY WITHOUT ANY OBSERVED DEFECTS. SAMPLES WERE FORWARDED TO MANUFACTURING (HOLDREGE) ON 25MAY2018 FOR FURTHER REVIEW. SECOND INVESTIGATION SUMMARY: ON 30MAY2018, HOLDREGE RECEIVED TEN (10) 0.5ML, 8MM, 30G SYRINGES IN OPENED POLYBAG FROM BATCH # 7100858. ALL SAMPLES WERE DECONTAMINATED PER HSTR-17 PRIOR TO BEING EVALUATED. UPON EVALUATION BY QE AH, SIMILAR FINDINGS TO THOSE DOCUMENTED DURING INITIAL INVESTIGATION PERFORMED AT BD FRANKLIN LAKES WERE NOTED. THE DAMAGED SAMPLE WAS INSPECTED FOR ANY DAMAGE TO THE HUB ASSEMBLY (HUB, CANNULA, SHIELD) AND NONE WAS NOTED. PROBABLE ROOT CAUSE IS LIKELY TO BE A JAM ON THE PREP DIAL DURING ASSEMBLY, RESULTING IN THE CRACKED BARREL AND ACCOUNTING FOR NO DAMAGE TO THE HUB ASSEMBLY. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 7100858. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE TWO (2) NOTIFICATIONS [200693429, 200691866] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. THERE WAS ONE (1) NOTIFICATION [200694301] NOTED FOR CRACKED HUBS. SEVERITY: S_2__; OCCURRENCE: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 1ST RELATED COMPLAINT FOR CLOG AND THE 1ST RELATED COMPLAINT FOR BROKEN BARREL ON LOT # 7100858. BASED ON THE SAMPLES / PHOTO(S) RECEIVED THE INVESTIGATION CONCLUDED: - CONFIRMED: BD WAS ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE (CRACKED BARREL). BASED ON THE ABOVE, NO CAPA IS REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 1

INITIAL REPORTER PHONE#: (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED WITH THE USE OF THE BD ULTRA-FINE INSULIN SYRINGE AND NEEDLE THERE WAS AN ISSUE WITH THE BARREL BEING BROKEN. IT WAS STATED ¿AFTER OPENING A NEW SYRINGE PACK, AT THE BEGINNING OF ASPIRATION THE INSULIN, THE CUSTOMER REALIZED THE SYRINGE DID NOT PULL THE MEDICINE. ONLY AIR WAS COMING AND WHEN SHE LOOKED IN THE SYRINGE SHE NOTICED THE CYLINDER WAS CRACKED.¿ THERE WAS NO REPORT OF INJURY OR FURTHER MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
355716 BD ULTRA-FINE INSULIN SYRINGE AND NEEDLE INSULIN SYRINGE AND NEEDLE FMF BD MEDICAL - DIABETES CARE 7100858

Patients

Seq Age Sex Outcome Treatment
1 Other