FDA Adverse Event Malfunction Summary report: N

PORTEX® SPINAL ANESTHESIA TRAYS

MDR report key: 7508213 · Received May 11, 2018

Report

Report Number
3012307300-2018-10245
Event Type
Malfunction
Date Received
May 11, 2018
Date of Event
January 17, 2017
Report Date
May 11, 2018
Manufacturer
SMITHS MEDICAL ASD, INC.
Product Code
CAZ
UDI-DI
20351688082333
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REVIEW OF DEVICE HISTORY RECORDS AND INCOMING RECORDS FOUND NO DISCREPANCIES OR ANOMALIES RELEVANT TO THE COMPLAINT. ASSAY TESTING, PERFORMED BY THE SUPPLIER PRIOR TO RELEASE OF BUPIVACAINE-MARCAINE LOT # 641003A, LIDOCAINE LOT # 63-384-EV AND EPINEPHRINE LOT # 631053A, WAS WITHIN SPECIFICATION. REVIEW OF THE STERILIZATION CERTIFICATE FOR THE TRAY LOT NUMBER 3241851 ((B)(4)) REVEALED THAT THE PRODUCT WAS PROCESSED ACCORDING TO VALIDATED SPECIFICATION REQUIREMENTS AND PARAMETERS. SMITH MEDICAL PERIODICALLY MONITORS SUPPLIED DRUG COMPONENTS STABILITY POST ETHYLENE OXIDE STERILIZATION THROUGHOUT THEIR EXPIRY. OUR RECORDS SHOW THAT THE SUPPLIED DRUGS COMPONENTS CONTINUE TO MEET POTENCY SPECIFICATIONS. COMPLAINT HISTORY REVIEW DOES NOT INDICATE ANY UPWARD TRENDS FOR THE DRUG LOT NUMBERS, REPORTED IN THIS COMPLAINT. THE ANESTHETICS ARE SUPPLIED ITEMS AND THE COMPLAINT NOTIFICATION WAS FORWARDED TO THE SUPPLIER. POSSIBLE CAUSES OF LACK OF EFFECT MAY BE ADMINISTRATION TECHNIQUE OR PATIENT ANATOMICAL VARIATIONS, INCLUDING PATHOLOGICAL OR PSYCHOLOGICAL

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MARCAINE CONTAINED IN THE PORTEX® SPINAL ANESTHESIA TRAY WAS INEFFECTIVE. THE PATIENT WAS PLACED UNDER GENERAL ANESTHESIA AS A RESULT. NO PATIENT DEATH OR SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
349244 PORTEX® SPINAL ANESTHESIA TRAYS ANESTHESIA CONDUCTION KIT CAZ SMITHS MEDICAL ASD, INC. 3241851 20351688082333

Patients

Seq Age Sex Outcome Treatment
1