FDA Adverse Event Malfunction Summary report: N

XPS® BLADE

MDR report key: 7500276 · Received May 9, 2018

Report

Report Number
3004209178-2018-10494
Event Type
Malfunction
Date Received
May 9, 2018
Date of Event
March 26, 2018
Report Date
May 9, 2018
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
EQJ
UDI-DI
20681490047443
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT ANALYSIS CONFIRMED THAT ONE DEVICE, PART NUMBER 1884002 FROM LOT HG1WYGP, WAS RECEIVED WITH THE INNER CUTTER TIP BROKEN OFF. A MICROSCOPE AND CALIPERS WERE USED TO FURTHER ASSESS THE RETURNED SAMPLE. THE FRAGMENTED PORTION WAS APPROXIMATELY 0.24" IN LENGTH WITH THE DETACHMENT SITE OCCURRING AT THE FIRST PROXIMAL TOOTH VALLEY. THE INNER ASSEMBLY WAS DEFORMED IN SUCH A WAY THAT INDICATES THE INNER BLADE TEETH IMPACTED THE OUTER CUTTING SURFACE WHILE MOVING BOTH THE CLOCKWISE AND COUNTERCLOCKWISE DIRECTIONS WHICH RESULTED IN THE OBSERVED BREAK. THE IMPACTS MARKINGS WERE AT THE SECOND AND THIRD PROXIMAL VALLEYS, AND THERE WAS UNEVEN WEAR AT THE CUTTING TIP WHICH MAY INDICATE AN INTERFERENCE FIT BETWEEN THE INNER AND OUTER ASSEMBLIES AT THE TIP. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE TIP OF THE BLADE BROKE OFF. THE TIP WAS RETRIEVED. THERE WAS A DELAY OF LESS THAN 10 MINUTES AND NO PATIENT INJURY OR IMPACT.