FDA Adverse Event Malfunction Summary report: N

*

MDR report key: 748957 · Received July 26, 2005

Report

Report Number
MW1036093
Event Type
Malfunction
Date Received
July 26, 2005
Date of Event
June 29, 2005
Report Date
June 30, 2005
Manufacturer
AIR GAS: DISTRIBUTOR VNG
Product Code
ECX
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
OH, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

HELIUM HIGH PURITY 4.8 GRADE 6,000 PSI TANK USED ON SEEDNET MACHINE FOR CRYOABLATION. DURING NEEDLE TEST PHASE, HELIUM IS SENT TO NEEDLES TO CHECK INTEGRITY, ICE FORMATION NOTED ON TIPS OF NEEDLES. DURING LAST 50 SECS OF TEST PHASE, HELIUM IS SENT TO TIPS OF NEEDLES TO THAW ICE FORMATION BUT, THIS DID NOT OCCUR. THE ICE FORMATION CONTINUED TO OCCUR ON NEEDLE TIPS. HELIUM TANK REMOVED FROM MACHINE AND ISOLATED. NO PATIENT INJURY OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 * HELIUM HIGH PUIRTY 4.8 GRADE ECX AIR GAS: DISTRIBUTOR VNG UN1046 *

Patients

Seq Age Sex Outcome Treatment
1 *