FDA Adverse Event Malfunction Summary report: N

ROTAFLOW CENTRIFUGAL PUMP SYSTEM

MDR report key: 7486963 · Received May 4, 2018

Report

Report Number
8010762-2018-00159
Event Type
Malfunction
Date Received
May 4, 2018
Date of Event
April 11, 2018
Report Date
February 22, 2019
Manufacturer
MAQUET CARDIOPULMONARY AG
Product Code
DTQ
PMA / PMN Number
K991864
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

(B)(4). A FOLLOW-UP MEDWATCH WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. REFERENCE EXEMPTION # E2018002. (B)(4).

Additional Manufacturer Narrative · 0

MAQUET MEDICAL SYSTEMS,USA(IMPORTER)SUBMITS THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET CARDIOPULMONARY AG KEHLER STRASSE 31, 76437 RASTATT, GERMANY. A FOLLOW-UP MEDWATCH WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. REFERENCE EXEMPTION # E2018002. IMPORTER- MAQUET MEDICAL SYSTEMS USA 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON- (B)(4). THE ROTAFLOW CONSOLE WAS INVESTIGATED WITH THE MAQUET - SERVICE ORDER NUMBER (B)(4) BY A GETINGE SERVICE TECHNICIAN ON THE 2018-04-26: A FAILURE COULD BE CONFIRMED THEREFORE THE CONTROL BOARD KIT (PCBA) WAS EXCHANGED. NO EXCHANGED PART DOES EXIST THEREFORE NO PROBABLE ROOT CAUSE IS POSSIBLE. NO SERVICE PROTOCOL DOES EXIST THEREFORE IT CANNOT BE GUARANTEED THAT THE MACHINE IS FULLY FUNCTIONAL AGAIN. THE DATA IS ALSO BEING HANDLED THROUGH A DESIGNATED MAQUET CARDIOPULMONARY TRENDING AND APPLICABLE INVESTIGATION PROCESS. DUE TO THIS NO FURTHER INVESTIGATION INITIATIONS WILL BE COMPLETED AT THIS TIME.

Description of Event or Problem · 0

(B)(4). DURING MAINTENANCE/ SERVICE A "HEAD ERROR" WAS SHOWN WHEN STARTING. NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

INTERNAL REFERENCE#:(B)(4). ONE SUPPORT#: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331649 ROTAFLOW CENTRIFUGAL PUMP SYSTEM CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS DTQ MAQUET CARDIOPULMONARY AG ROTAFLOW CONSOLE

Patients

Seq Age Sex Outcome Treatment
1