HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0
Report
- Report Number
- 3007042319-2018-01734
- Event Type
- Malfunction
- Date Received
- May 3, 2018
- Date of Event
- April 3, 2018
- Report Date
- September 28, 2018
- Manufacturer
- HEARTWARE, INC.
- Product Code
- DSQ
- UDI-DI
- 00888707000420
- PMA / PMN Number
- P100047
- Removal / Correction Number
- Z-1903-2018
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
PRODUCT EVENT SUMMARY: THE CONTROLLER (CON307103) AND FOUR BATTERIES (BAT554165, BAT554168, BAT554256, BAT554261) WERE RETURNED FOR EVALUATION. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICES IN RELATION TO THE REPORTED EVENT. FAILURE ANALYSIS OF THE RETURNED BATTERIES REVEALED THAT THE DEVICES PASSED FUNCTIONAL TESTING AND VISUAL EXAMINATION. FAILURE ANALYSIS OF THE RETURNED CONTROLLER REVEALED THAT THE DEVICE PASSED FUNCTIONAL TESTING. VISUAL INSPECTION OF THE CONTROLLER UNDER 10X MAGNIFICATION REVEALED A HAIRLINE CRACK AROUND POWER PORT TWO (2). AN INTERNAL INSPECTION DID NOT REVEAL EVIDENCE OF FLUID INGRESS. THE OBSERVED HAIRLINE CRACK WAS NOT RELATED TO THE REPORTED EVENT. BASED ON AN INTERNAL INVESTIGATION, THE ROOT CAUSE OF THE HAIRLINE CRACK WAS DETERMINED TO BE DUE TO CHEMICAL ADDITIVES APPLIED TO THE POWER PORT GASKETS DURING THE MANUFACTURING PROCESS. THE CHEMICAL ADDITIVES CONTRIBUTED TO ENVIRONMENTAL STRESS CRACKING. LOG FILE ANALYSIS REVEALED THAT THE CONTROLLER CONTAINED A FEATURE THAT RECORDS WHETHER A POWER SOURCE EXPERIENCED A COMMUNICATION ERROR OR A DISCONNECTION WITHIN EACH 15-MINUTE INTERVAL. ANALYSIS OF THE ALARM LOG FILE REVEALED FOUR (4) CRITICAL BATTERY ALARMS DUE TO COMMUNICATION ERRORS INVOLVING BAT554261 AND BAT554168. ANALYSIS OF THE DATA LOG FILE ALSO REVEALED SEVERAL MOMENTARY DISCONNECTIONS INVOLVING BAT554261. MOMENTARY DISCONNECTIONS WILL RESULT IN AN AUDIBLE TONE OR "BEEP." AS A RESULT, THE REPORTED CRITICAL BATTERY ALARM AND BEEPING EVENTS WERE CONFIRMED. THE MOST LIKELY ROOT CAUSE OF THE REPORTED CRITICAL BATTERY ALARM EVENT CAN BE ATTRIBUTED TO COMMUNICATION ERRORS BETWEEN THE CONTROLLER AND BATTERIES. THE MOST LIKELY ROOT CAUSE OF THE REPORTED BEEPING CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS BETWEEN THE CONTROLLER AND BATTERIES. AN INTERNAL INVESTIGATION IS OPEN TO EVALUATE MOMENTARY DISCONNECTIONS. ADDITIONAL PRODUCTS: BATTERY BAT554168. RETURN DATE: (B)(6)2018. FDA METHOD CODE(S): 10, 4112. FDA RESULTS CODE(S): 3213 FDA CONCLUSION CODE(S): 4307 BATTERY BAT554261. YES, RETURN DATE: (B)(6) 2018. FDA METHOD CODE(S): 10, 4112 FDA RESULTS CODE(S): 3213. FDA CONCLUSION CODE(S): 4307, 12 BATTERY BAT554165 YES, RETURN DATE: 2018-07-16. FDA METHOD CODE(S): 10, 4112. FDA RESULTS CODE(S): 213. FDA CONCLUSION CODE(S): 67 BATTERY BAT554256 D10: YES, RETURN DATE: (B)(6) 2018. FDA METHOD CODE(S): 10, 4112. FDA RESULTS CODE(S): 213. FDA CONCLUSION CODE(S): 67 MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. CONTINUATION OF MEDICAL DEVICES: 6935M62 LEAD IMPLANTED (B)(6) 2016 OTHER DEVICES INVOLVED IN THIS EVENT: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY/ (B)(4)/ MODEL #: 1650 / EXPIRATION DATE: 2018-04-30 UDI #: (B)(4) MFG DATE: 2017-04-30 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650 / EXPIRATION DATE: 2018-04-30 UDI #: (B)(4) MFG DATE: 2017-04-30 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY/ (B)(4)/ MODEL #: 1650 / EXPIRATION DATE: 2018-04-30 UDI #: (B)(4) MFG DATE: 2017-04-30 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650 / EXPIRATION DATE: 2018-04-30 UDI #:(B)(4) MFG DATE: 2017-04-30 (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE PATIENT EXPERIENCED CRITICAL BATTERY ALARMS AND RANDOM BEEPS WHILE ON THE ATTACHED CONTROLLER WITH THE ATTACHED BATTERIES. THE BATTERY CAPACITIES WERE ABOVE 10% WHEN THE CRITICAL BATTERY ALARMS OCCURRED. THE BATTERIES AND CONTROLLER WERE EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 325852 | HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0 | VENTRICULAR (ASSISST) BYPASS | DSQ | HEARTWARE, INC. | 1420 | 00888707000420 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | DTBA1QQ ICD, 459888 LEAD, 5076-52 LEAD |