FDA Adverse Event Death Summary report: N

NEWLIFE ELITE

MDR report key: 7484593 · Received May 3, 2018

Report

Report Number
3004972304-2018-00012
Event Type
Death
Date Received
May 3, 2018
Report Date
October 17, 2018
Manufacturer
CAIRE INC.
Product Code
CAW
PMA / PMN Number
K872534
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

"PURSUANT TO TITLE 21 - FOOD AND DRUGS, CHAPTER I - FOOD AND DRUG ADMINISTRATION DEPARTMENT OF HEALTH AND HUMAN SERVICES, SUBCHAPTER H -0 MEDICAL DEVICE, PART 803 - MEDICAL DEVICE REPORTING, SUBPART A - GENERAL PROVISIONS, SECTION 803.16, NEITHER THIS REPORT NOR ANY INFORMATION SUBMITTED HEREIN CONSTITUTES AN ADMISSION BY CAIRE INC. THAT THE DEVICE STATED IN THIS REPORT, CAIRE INC., OR CAIRE INC.'S EMPLOYEES, CAUSED OR CONTRIBUTED TO THE REPORTABLE EVENT STATED HEREIN." THE UNIT WAS RETURNED FOR AN EVALUATION. THE DEVICE DOES NOT OPERATE AS INTENDED DUE TO MISSING COMPONENTS, LEAKS, AND HEAVY CONTAMINATION. THE DEVICE REQUIRES REPAIR AND MAINTENANCE.

Description of Event or Problem · 0

THIS REPORT WAS ORIGINALLY SUBMITTED ON 10/17/2018, AND IS BEING RESUBMITTED ON 4/8/2020 AS THE ORIGINAL SUBMISSION FAILED TO GO THROUGH. THE OXYGEN CONCENTRATOR HAD BEEN WORKING CORRECTLY, AND WAS GIVEN ITS PERIODIC SERVICE ON MAY 18TH 2017. IT APPEARS TO HAVE STOPPED WORKING AFTERWARDS. THE DEVICE USER WAS FOUND IN A DETERIORATED STATE, AND ACCORDING TO HER HUSBAND, NO GAS WAS COMING OUT OF THE MASK ATTACHED TO THE CONCENTRATOR. SADLY, THE DEVICE USER PASSED AWAY ON (B)(6) 2017.

Additional Manufacturer Narrative · 1

WE ARE ATTEMPTING TO GET THE UNIT BACK FOR AN EVALUATION. IF ANY NEW INFORMATION IS DISCOVERED, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE OXYGEN CONCENTRATOR HAD BEEN WORKING CORRECTLY, AND WAS GIVEN ITS PERIODIC SERVICE ON (B)(6) 2017. IT APPEARS TO HAVE STOPPED WORKING AFTERWARDS. THE DEVICE USER WAS FOUND IN A DETERIORATED STATE, AND ACCORDING TO HER HUSBAND, NO GAS WAS COMING OUT OF THE MASK ATTACHED TO THE CONCENTRATOR. SADLY, THE DEVICE USER PASSED AWAY ON (B)(6) 2017.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
327974 NEWLIFE ELITE CONCENTRATOR, OXYGEN, STATIONARY CAW CAIRE INC. AS005-231

Patients

Seq Age Sex Outcome Treatment
1 Death