FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES

MDR report key: 7479944 · Received May 2, 2018

Report

Report Number
1917413-2018-01228
Event Type
Malfunction
Date Received
May 2, 2018
Date of Event
May 22, 2017
Report Date
September 5, 2018
Manufacturer
BECTON, DICKINSON & CO.
Product Code
JKA
UDI-DI
50382903679602
PMA / PMN Number
K945952
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: THE DEVICE HISTORY RECORDS WERE REVIEWED FOR BATCH 7033697 WITH NO ISSUES BEING IDENTIFIED. BASED ON THE INVESTIGATION, THE MOST LIKELY ROOT CAUSE IS THAT AIR BUBBLES IN GEL ARE CAUSED BY AIR POCKETS IN THE GEL MATERIAL OR BY AIR IN THE LINES DURING THE DISPENSE PROCESS. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. THE BD PAS BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES HAD AIR BUBBLES IN THE GEL.

Additional Manufacturer Narrative · 1

BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE SAMPLES WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR 91882 WITH THE INCIDENT LOT WAS OBSERVED. A REVIEW OF THE MANUFACTURING RECORDS WAS COMPLETED FOR THE INCIDENT LOT AND NO ISSUES WERE IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES HAD AIR BUBBLES IN THE GEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
324522 BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. 7033697 50382903679602

Patients

Seq Age Sex Outcome Treatment
1 Other