BD VACUTAINER® K3E 7.2MG PLUS BLOOD COLLECTION TUBES
Report
- Report Number
- 9617032-2018-00103
- Event Type
- Malfunction
- Date Received
- May 1, 2018
- Date of Event
- April 9, 2018
- Report Date
- April 11, 2018
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- PJE
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INVESTIGATION RESULTS: SUMMARY: BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE SAMPLES WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR UNDERFILL WITH THE INCIDENT LOT WAS NOT OBSERVED. EVALUATION AND TESTING OF THE CUSTOMER SAMPLES WAS PERFORMED AND UNDERFILL WAS NOT OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES OF THIS ISSUE ARE IDENTIFIED. CONCLUSION: BASED ON EVALUATION OF THE CUSTOMER SAMPLES, THE CUSTOMER¿S INDICATED FAILURE MODE FOR UNDERFILL WITH THE INCIDENT LOT WAS NOT OBSERVED. FURTHER INVESTIGATION HAS BEEN INITIATED THROUGH CAPA 161123. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES ARE IDENTIFIED. ROOT CAUSE: CAPA 161123 HAS BEEN INITIATED TO DOCUMENT FURTHER INVESTIGATION AND ROOT CAUSE ANALYSIS RELATING TO THIS ISSUE. THE INVESTIGATION IS CURRENTLY ON-GOING AND WILL BE UPDATED AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED. RATIONALE: BASED ON AN ASSESSMENT OF SEVERITY AND FREQUENCY, IT WAS DETERMINED THAT A CAPA IS REQUIRED AT THIS TIME IN ORDER TO DETERMINE THE ROOT CAUSE ASSOCIATED WITH THIS ISSUE AND IMPLEMENT CORRECTIVE ACTIONS. THE INVESTIGATION IS CURRENTLY ON-GOING AND FURTHER IMPROVEMENTS WILL BE MADE AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED.
IT WAS REPORTED THAT SEVERAL EXTERNAL PRODUCERS FOUND UNDERFILLING WHEN USING BD VACUTAINER® K3E 7.2MG PLUS BLOOD COLLECTION TUBES. IN ADDITION, IT WAS REPORTED THE LABORATORY HAS FOUND THE TUBES TO BE FILLING SLOWER THAN OTHER TUBES. THERE WAS NO REPORT OF EXPOSURE, SERIOUS INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 318447 | BD VACUTAINER® K3E 7.2MG PLUS BLOOD COLLECTION TUBES | BLOOD COLLECTION TUBE | PJE | BECTON, DICKINSON AND COMPANY (BD) | 7114655 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |