VARIS VISION TREATMENT EVI
Report
- Report Number
- 2914292-2005-00008
- Event Type
- Other
- Date Received
- October 31, 2005
- Date of Event
- May 26, 2005
- Report Date
- October 30, 2005
- Manufacturer
- VARIAN MEDICAL SYSTEMS
- Product Code
- IYE
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- INVALID DATA
Narratives
DURING THE INVESTIGATION PERFORMED BY VARIAN IT WAS DISCOVERED THAT IF A SYMMETRIC FIELD IS MODED UP WHEN THE LINAC ACTUAL POSITIONS ARE ASYMMETRIC, AND THE LINAC X1 OR Y1 JAW VALUE IS 1/2 OF THE PLAN SYMMETRIC VALUE, THEN THE X2 OR Y2 JAW VALUE CAN BE OUT OF POSITION AND VARIS VISION WILL NOT INHIBIT TREATMENT. IF THE OPERATOR OVERLOOKS THE FIELD SIZE DISCREPANCY, THE PT COULD BE TREATED WITH THE WRONG FIELD. UPON INITIAL REVIEW OF THIS ISSUE ON JUNE 14TH, 2005, THE HAZARD ANALYSIS TEAM INDICATED THAT A SERIOUS INJURY WAS NOT LIKELY. HOWEVER, UPON FURTHER REVIEW ON JUNE 21, 2005, IT WAS DETERMINED THAT THIS ISSUE COULD LIKELY RESULT IN A SERIOUS INJURY TO A CRITICAL STRUCTURE IF IT WERE TO RECCUR. A PRODUCT NOTIFICATION LETTER WAS DISTRIBUTED ON SEPTEMBER 20, 2005 TO ALL RELEVANT CUSTOMERS WARNING THEM OF THIS ISSUE. A CORRECTION NOTICE WAS SENT TO THE LOCAL RECALL COORDINATOR ON OCTOBER 4, 2005. A FIX FOR THIS PROBLEM HAS BEEN CREATED AND WILL BE APPLIED TO THE CUSTOMER SYSTEMS.
A CUSTOMER REPORTED THE FOLLOWING EVENT: THE PLANNED FIELD SIZE WAS 20X20. THE Y ACTUAL VALUE WAS 20, BUT THE X ACTUAL WAS 11. VARIS DID NOT SHOW AN ORANGE BOX ON THE X SIZE TO SHOW IT WAS OUT OF TOLERANCE. THE "RADIOGS" USED ASU AND THE LINAC WENT TO THE PLANNED POSITION. THE CUSTOMER NOTICED THE DISCREPANCY, THEREFORE THERE WAS NOT A MISTREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VARIS VISION TREATMENT EVI | ACCELERATOR, LINEAR, MEDICAL-SOFTWARE | IYE | VARIAN MEDICAL SYSTEMS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |