FDA Adverse Event Malfunction Summary report: N

ADVANCED PERSONAL DOUBLE BREASTPUMP (WITHOUT BATTERY PACK)

MDR report key: 7472097 · Received April 30, 2018

Report

Report Number
1419937-2018-00114
Event Type
Malfunction
Date Received
April 30, 2018
Date of Event
April 18, 2018
Report Date
April 30, 2018
Manufacturer
MEDELA LLC
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT POWER SUPPLY WAS SENT TO THE CUSTOMER. THE ISSUE WITH A DAMAGED REV N POWER SUPPLY FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4), WHICH WAS TRENDED AS PART OF THE EFFECTIVENESS CHECK FOR (B)(4), WHICH FOUND THAT THE POWER SUPPLIES WERE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. AS A PART OF ROUTINE CONTINUOUS IMPROVEMENT ACTIVITIES, THE REV N POWER SUPPLY WAS REPLACED WITH A REV P POWER SUPPLY, MANUFACTURED UNDER A REVISED DESIGN AND BY A DIFFERENT MANUFACTURER.

Description of Event or Problem · 1

ON (B)(6) 2018, THE CUSTOMER ALLEGED TO MEDELA LLC THAT THE HOUSING ON THE POWER SUPPLY FOR HER PUMP IN STYLE BREAST PUMP CRACKED AND THE METAL PRONGS ARE BROKEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
317283 ADVANCED PERSONAL DOUBLE BREASTPUMP (WITHOUT BATTERY PACK) PUMP, BREAST, POWERED HGX MEDELA LLC 57018W 9207010 REV N

Patients

Seq Age Sex Outcome Treatment
1