FDA Adverse Event Other Summary report: N

OP / EMS

MDR report key: 747178 · Received November 1, 2005

Report

Report Number
MW1037070
Event Type
Other
Date Received
November 1, 2005
Report Date
November 1, 2005
Manufacturer
ALLEGIANCE EMS
Product Code
LRW
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
TX, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

DURING LAPAROSCOPIC CHOLECYSTECTOMY, THE INSULATION ON "SHEERS" GAVE THE PATIENT A 1.5 CM BURN ON MUSCLE TISSUE AND LEFT A SMALL BURN ON TIP OF TROCAR IT FED THROUGH. THE INSULATION DID APPEAR TO BE INTACT. IT DID HAVE A SLEEVE OVER THE AREA LIKE A PREVIOUS REPAIR, THIS SLEEVE BURNED AND WAS MELTED APPEARING. NO PATIENT TREATMENT WAS NEEDED AS BURNED AREA WAS SAME TYPE OF TISSUE THAT WOULD HAVE BEEN AFFECTED HAD A BLEEDER BEEN BOVIED. THE BURNED AREA WAS IN MUSCLE AREA AND DID NOT AFFECT CRUCIAL AREAS SUCH AS DIAPHRAGM OR BOWEL. EQUIPMENT AND INSTRUMENTS ARE VENDOR OWNED AND MAINTAINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OP / EMS SHEARS - LAPAROSCOPIC SCISSORS LRW ALLEGIANCE EMS * *

Patients

Seq Age Sex Outcome Treatment
1 47 YR Other SHEARS - PART OF LAPAROSCOPY GENERAL SET.