FDA Adverse Event
Other
Summary report: N
OP / EMS
MDR report key: 747178
·
Received November 1, 2005
Report
- Report Number
- MW1037070
- Event Type
- Other
- Date Received
- November 1, 2005
- Report Date
- November 1, 2005
- Manufacturer
- ALLEGIANCE EMS
- Product Code
- LRW
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- TX, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
DURING LAPAROSCOPIC CHOLECYSTECTOMY, THE INSULATION ON "SHEERS" GAVE THE PATIENT A 1.5 CM BURN ON MUSCLE TISSUE AND LEFT A SMALL BURN ON TIP OF TROCAR IT FED THROUGH. THE INSULATION DID APPEAR TO BE INTACT. IT DID HAVE A SLEEVE OVER THE AREA LIKE A PREVIOUS REPAIR, THIS SLEEVE BURNED AND WAS MELTED APPEARING. NO PATIENT TREATMENT WAS NEEDED AS BURNED AREA WAS SAME TYPE OF TISSUE THAT WOULD HAVE BEEN AFFECTED HAD A BLEEDER BEEN BOVIED. THE BURNED AREA WAS IN MUSCLE AREA AND DID NOT AFFECT CRUCIAL AREAS SUCH AS DIAPHRAGM OR BOWEL. EQUIPMENT AND INSTRUMENTS ARE VENDOR OWNED AND MAINTAINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OP / EMS | SHEARS - LAPAROSCOPIC SCISSORS | LRW | ALLEGIANCE EMS | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Other | SHEARS - PART OF LAPAROSCOPY GENERAL SET. |