FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY

MDR report key: 7469784 · Received April 27, 2018

Report

Report Number
1213809-2018-00268
Event Type
Malfunction
Date Received
April 27, 2018
Date of Event
April 6, 2018
Report Date
June 29, 2018
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903097013
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: NO SAMPLES INCLUDING PHOTOS WERE PROVIDED FOR EVALUATION. HOWEVER THE COMPLAINT IS FOR SHORT COUNT, THEREFORE, IT IS CONFIRMED BASED ON DESCRIPTION. DHR REVIEW FOR BATCH 7240832 (P/N 309701): MANUFACTURING DATES: 09/29/2017 TO 10/01/2017. BATCH QUANTITY WAS (B)(4). ALL VISUAL INSPECTIONS WERE PERFORMED AS PER REQUIREMENT WITH NO QUALITY NOTIFICATIONS RELATED TO THE COMPLAINT DEFECT. BATCH 7240832 WAS INSPECTED AND ACCEPTED BASED ON MEETING OUR INSPECTION CONTROL PLAN AND SUBSEQUENTLY APPROVED FOR SHIPMENT. INVESTIGATION CONCLUSION: MACHINE LOGS REVEAL ISSUES WITH SYRINGE LOADING ROBOT RESULTING IN LOW COUNT TRAYS. THE ISSUE WAS ADDRESSED BEFORE PRODUCTION RESUMED. HOWEVER IT IS POSSIBLE A LIMITED NUMBER OF TRAYS ESCAPED DETECTION.

Description of Event or Problem · 0

IT WAS REPORTED WITH THE USE OF THE BD LUER-LOK¿ SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY THERE WAS AN ISSUE WITH THE STOPPER BEING MISALIGNED. IT WAS STATED ¿SEVERAL OF THE BOXES WITHIN THESE CASES ONLY HAVE 13-19 SYRINGES IN THE PACKAGE VERSUS 25 THAT SHOULD BE IN THE BOX. ADDITIONALLY, THE SYRINGES ARE NOT PROPERLY ALIGNED.¿ THERE WAS NO REPORT OF INJURY OR FURTHER MEDICAL INTERVENTION.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED WITH THE USE OF THE BD LUER-LOK¿ SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY THERE WAS AN ISSUE WITH THE STOPPER BEING MISALIGNED. IT WAS STATED ¿SEVERAL OF THE BOXES WITHIN THESE CASES ONLY HAVE 13-19 SYRINGES IN THE PACKAGE VERSUS 25 THAT SHOULD BE IN THE BOX. ADDITIONALLY, THE SYRINGES ARE NOT PROPERLY ALIGNED.¿ THERE WAS NO REPORT OF INJURY OR FURTHER MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
313640 BD LUER-LOK¿ SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 7240832 30382903097013

Patients

Seq Age Sex Outcome Treatment
1 Other