FDA Adverse Event Malfunction Summary report: N

TRUEMETRIX

MDR report key: 7468695 · Received April 27, 2018

Report

Report Number
1000113657-2018-00490
Event Type
Malfunction
Date Received
April 27, 2018
Date of Event
April 9, 2018
Report Date
April 27, 2018
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00021292006075
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. TEST STRIPS NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE : MLC-55 USER HAD AN INACCURATE REFERENCE: COMPETITOR'S METER: THE END USER IS COMPARING RESULTS OBTAINED FROM TRIVIDIA'S BGM SYSTEM TO THE RESULTS FROM A COMPETITOR'S BGM SYSTEM TEST STRIP UDI# (B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER ON (B)(4) 2018 IN A FOLLOW-UP CALL IN ORDER TO ENSURE THAT THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN; CUSTOMER STATED THAT THE REPLACEMENT PRODUCT IS WORKING TO THEIR SATISFACTION AND THEY HAVE NOT HAD ANY MEDICAL INTERVENTION SINCE THE LAST CALL.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR LOW BLOOD GLUCOSE TEST RESULTS. SON IS CALLING ON BEHALF OF THE CUSTOMER. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 130, 132, 170 AND 149 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 200 - 250 MG/DL. SON STATED THAT HE IS CONCERNED TRUEMETRIX METER IS NOT GIVING CORRECT RESULTS DUE TO BEING 100 POINTS LOWER THAN ANOTHER DEVICE; ACTUAL METER TO METER COMPARISON RESULTS WERE NOT PROVIDED. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2018, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER; CUSTOMER WAS NOT PRESENT AT TIME OF CALL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 02/17/2019 AND OPEN VIAL DATE IS (B)(6) 2018. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
312095 TRUEMETRIX BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUEMETRIX MU2589 00021292006075

Patients

Seq Age Sex Outcome Treatment
1 0 YR