FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG

MDR report key: 7466307 · Received April 26, 2018

Report

Report Number
1917413-2018-00798
Event Type
Malfunction
Date Received
April 26, 2018
Date of Event
January 16, 2017
Report Date
June 13, 2018
Manufacturer
BECTON, DICKINSON & CO.
Product Code
JKA
UDI-DI
50382903686617
PMA / PMN Number
K972075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

OTHER DETAILS IS CORRECTED TO: SEE EVENT DESCRIPTION BELOW. DESCRIBE EVENT OR PROBLEM IS CORRECTED TO:IT WAS REPORTED THAT BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG HAD CLOTTING ISSUES. NO SERIOUS INJURY OR MEDICAL INTERVENTION. MEDICAL DEVICE BRAND NAME CORRECTED TO: BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG. UNIQUE IDENTIFIER (UDI) # CORRECTED TO: (B)(4). PMA / 510(K)# CORRECTED TO: K972075.

Additional Manufacturer Narrative · 0

MEDICAL DEVICE CATALOG # CORRECTED 362788. MEDICAL DEVICE LOT # CORRECTED 6173860. MEDICAL DEVICE EXPIRATION DATE CORRECTED 006/30/2017. DATE RECEIVED BY MANUFACTURER CORRECTED 01/16/2017.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG HAD CLOTTING ISSUES. NO SERIOUS INJURY OR MEDICAL INTERVENTION.

Additional Manufacturer Narrative · 1

BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FACILITY FOR EVALUATION; THEREFORE, THE INVESTIGATION WAS LIMITED. RETENTION SAMPLES WERE SELECTED FROM BD INVENTORY FOR EVALUATION, AND THE CUSTOMER'S INDICATED FAILURE MODE FOR (B)(6) WAS NOT OBSERVED AS ALL SAMPLES MET SPECIFICATIONS. A REVIEW OF THE MANUFACTURING RECORDS WAS COMPLETED FOR THE INCIDENT LOT AND NO ISSUES WERE IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD VACUTAINER® K2 EDTA (K2E) 10.8 MG BLOOD COLLECTION TUBES HAD CLOTTING IN TUBES

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310087 BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. 6173860 50382903686617

Patients

Seq Age Sex Outcome Treatment
1 Other