SPYSCOPE DS ACCESS & DELIVERY CATHETER
Report
- Report Number
- 3005099803-2018-01263
- Event Type
- Malfunction
- Date Received
- April 26, 2018
- Date of Event
- April 3, 2018
- Report Date
- April 3, 2018
- Manufacturer
- BOSTON SCIENTIFIC - SPENCER
- Product Code
- FBN
- UDI-DI
- 08714729863236
- PMA / PMN Number
- K142922
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
A VISUAL ASSESSMENT WAS PERFORMED AFTER DISINFECTION. AS RECEIVED, THE WORKING CHANNEL SLEEVE PROTRUDED, CONFIRMING THE COMPLAINT. A FUNCTIONAL ASSESSMENT WAS PERFORMED. MAXIMUM WORKING CHANNEL SLEEVE PROTRUSION WAS OBSERVED WHEN THE DISTAL TIP WAS ARTICULATED BY TURNING THE LARGE KNOB IN THE COUNTERCLOCKWISE DIRECTION. THE DISTAL TIP WAS CUT. THE DISTAL CAP WAS REMOVED. THE CATHETER WAS CUT OPEN USING THE CUTTING FIXTURE. THE WORKING CHANNEL SLEEVE WAS REMOVED. WITNESS MARKS WERE NOTED ON THE PEBAX. THE WHITE AREAS ALONG BOND A APPEAR TO SHOW EVIDENCE OF ADHESION. THE COMPLAINT WAS CONSISTENT WITH THE REPORTED EVENT OF WORKING CHANNEL SLEEVE PROTRUDING. BASED ON THE INVESTIGATION RESULTS, THE UNDERLYING CAUSE OF WORKING CHANNEL SLEEVE PROTRUSION IS AN INSUFFICIENT BOND, PARTICULARLY THE SECOND HEAT CYCLE OF THE WORKING CHANNEL SLEEVE BONDING PROCESS [BOND B]. THEREFORE, THE COMPLAINT INVESTIGATION CONCLUSION CODE SELECTED FOR THE WORKING CHANNEL SLEEVE PROTRUSION ISSUE IS MANUFACTURING PROCESS DESIGN. AN INVESTIGATION IS UNDERWAY TO ADDRESS THIS ISSUE. A DHR (DEVICE HISTORY RECORD) REVIEW WAS PERFORMED AND NO DEVIATION WAS FOUND.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SPYSCOPE DIGITAL ACCESS AND DELIVERY CATHETER WAS USED IN THE BILE DUCT DURING A CHOLANGIOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THIS DEVICE WAS INSERTED INTO THE BILE DUCT WITH THE AID OF A GUIDEWIRE. THE GUIDEWIRE WAS THEN REMOVED. WHILE OBSERVING THE TUMOR, IT WAS NOTICED THAT THE WORKING CHANNEL SLEEVE OF THE SPYSCOPE DS PROTRUDED. REPORTEDLY, THERE WERE NO ACCESSORY DEVICE INSIDE THE SPYSCOPE DS WHEN THE WORKING CHANNEL SLEEVE PROTRUDED AND NO PART OF THE SPYSCOPE DS DEVICE WAS DETACHED. THE PROCEDURE WAS STILL COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
PATIENT'S EXACT AGE IS UNKNOWN; HOWEVER IT WAS REPORTED THAT THE PATIENT WAS OVER THE AGE OF 18. ALTHOUGH THE SUSPECT DEVICE HAS BEEN RECEIVED, THE EVALUATION HAS NOT BEEN COMPLETED. THEREFORE, THE CAUSE OF THE REPORTED MALFUNCTION HAS NOT BEEN DETERMINED. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MDR WILL BE FILED.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SPYSCOPE DIGITAL ACCESS AND DELIVERY CATHETER WAS USED IN THE BILE DUCT DURING A CHOLANGIOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THIS DEVICE WAS INSERTED INTO THE BILE DUCT WITH THE AID OF A GUIDEWIRE. THE GUIDEWIRE WAS THEN REMOVED. WHILE OBSERVING THE TUMOR, IT WAS NOTICED THAT THE WORKING CHANNEL SLEEVE OF THE SPYSCOPE DS PROTRUDED. REPORTEDLY, THERE WERE NO ACCESSORY DEVICE INSIDE THE SPYSCOPE DS WHEN THE WORKING CHANNEL SLEEVE PROTRUDED AND NO PART OF THE SPYSCOPE DS DEVICE WAS DETACHED. THE PROCEDURE WAS STILL COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 306576 | SPYSCOPE DS ACCESS & DELIVERY CATHETER | CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID | FBN | BOSTON SCIENTIFIC - SPENCER | M00546600 | 21698122 | 08714729863236 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |