FDA Adverse Event Malfunction Summary report: N

METRIX AIR

MDR report key: 7459236 · Received April 25, 2018

Report

Report Number
1000113657-2018-00480
Event Type
Malfunction
Date Received
April 25, 2018
Date of Event
April 5, 2018
Report Date
April 25, 2018
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00021292007010
PMA / PMN Number
K150052
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. TEST STRIPS NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-20-USER'S TEST STRIP HAD POOR STORAGE(KITCHEN). (B)(4). MANUFACTURER CONTACTED CUSTOMER ON (B)(6) 2018 IN A FOLLOW-UP CALL IN ORDER TO ENSURE THAT THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN; CUSTOMER STATED THAT THE REPLACEMENT PRODUCT IS WORKING TO THEIR SATISFACTION AND THEY HAVE NOT HAD ANY MEDICAL INTERVENTION SINCE THE LAST CALL.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 213, 175, 142, 147 AND 159 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100 - 120 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2018, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 10/27/2018 AND OPEN VIAL DATE IS FOUR DAYS. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1:213MG/DL DATE: (B)(6) 2018 TIME:11:15PM FASTING; RESULT 2:175MG/DL DATE: (B)(6) 2018 TIME:7:55PM FASTING; RESULT 3:142MG/DL DATE: (B)(6) 2018 TIME:9:54AM FASTING; RESULT 4:147MG/DL DATE: (B)(6) 2018 TIME:1:38PM FASTING; RESULT 5:159MG/DL DATE: (B)(6) 2019 TIME:8:38PM FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303587 METRIX AIR BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. METRIX AIR MU2446 00021292007010

Patients

Seq Age Sex Outcome Treatment
1 0 YR