FDA Adverse Event Malfunction Summary report: N

SERVO-I

MDR report key: 7458308 · Received April 25, 2018

Report

Report Number
8010042-2018-00204
Event Type
Malfunction
Date Received
April 25, 2018
Date of Event
March 28, 2018
Report Date
July 6, 2018
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
CBK
PMA / PMN Number
K123149
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).

Additional Manufacturer Narrative · 0

THE VENTILATOR WAS INVESTIGATED ON SITE AND AS A PREVENTIVE ACTION THE O2-SENSOR AND THE CONTROL PRINTED CIRCUIT BOARD (PCB) AND RETURNED FOR INVESTIGATION. THE O2-SENSOR MEASURES THE OXYGEN CONCENTRATION TO THE PATIENT. THE CONTROL PCB IS REGULATING THE BREATHING PATTERNS FOR DIFFERENT VENTILATION MODES. THE REPORTED ALARM FOR LOW O2-CONCENTRATION WAS NOT REPRODUCED DURING SIMULATED USE TEST OF THE RETURNED O2-SENSOR AND THE CONTROL PCB IN A REFERENCE VENTILATOR. THE RECEIVED DEVICE LOGS ARE NOT CONFIRMING THE REPORTED EVENT. THE REPORTED PROBLEM COULD NOT BE REPRODUCED, THEREFORE THE CAUSE COULD NOT BE DETERMINED IN THIS INVESTIGATION. GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE CONNECTED TO A PATIENT, THE VENTILATOR GENERATED ALARMS FOR LOW O2 CONCENTRATION. THERE WAS NO PATIENT HARM. (B)(4).

Description of Event or Problem · 0

IMPORTER REF. #: (B)(4). MANUFACTURER REF. #: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
305675 SERVO-I VENTILATOR, CONTINUOUS, FACILITY USE CBK MAQUET CRITICAL CARE AB

Patients

Seq Age Sex Outcome Treatment
1