ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2018-07938
- Event Type
- Malfunction
- Date Received
- April 25, 2018
- Report Date
- March 31, 2018
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER CAREGIVERS
Narratives
OF THE 56 ALLEGATIONS OF A MALFUNCTION, 43 PUMPS WERE RETURNED TO ANIMAS AND INVESTIGATED. OF THE INVESTIGATED PUMPS, 7 WERE FOUND TO BE MALFUNCTIONING DUE TO A SINGLE COMPONENT FAILURE. THIS REPORT IS PROCESSED UNDER EXEMPTION NUMBER E2015053. THIS MDR REPORT IS PART OF THE 227 PILOT PROGRAM.
THIS REPORT SUMMARIZES 56 MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT THE ONE TOUCH PING MODEL PUMP EXPERIENCED A CS 052/053/054/055 SLEEP ISSUE. THESE REPORTS WERE RECEIVED FROM VARIOUS SOURCES. THE PATIENTS ASSOCIATED WITH THIS ALLEGATION RANGED FROM 10-74 YEARS OF AGE AND BETWEEN 55 AND 228 POUNDS. OF THE REPORTED PATIENTS, 29 WERE MALE, 26 WERE FEMALE, AND 1 WAS UNKNOWN.
DEVICE EVALUATION: IN QUARTER 2 OF 2018, THERE WERE 1 INSTANCES OF CS 052/053/054/055 SLEEP FAILURE FOUND DURING INVESTIGATION. THIS REPORT IS PROCESSED UNDER EXEMPTION NUMBER (B)(4). THIS MDR REPORT IS PART OF THE (B)(4) PROGRAM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 305511 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |