FC500-MPL WITH MXP SOFTWARE
Report
- Report Number
- 1061932-2005-00023
- Event Type
- Other
- Date Received
- June 28, 2005
- Date of Event
- May 31, 2005
- Report Date
- June 28, 2005
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- LXG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
THE PROBLEM DESCRIBED IN THIS EVENT OCCURRED DURING THE INSTRUMENT INSTALLATION AND DID NOT AFFECT USER OR PATIENT. DUE TO THE POTENTIAL OF THIS EVENT TO RECUR UNKNOWINGLY FOR CD34+ PARAMETER, THERE IS A POSSIBILITY THAT TREATMENT DECISIONS MAY BE AFFECTED. A MALFUCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
A PROBLEM WAS ENCOUNTED WITH THE FC500-MPL (WITH MXP SOFTWARE) INSTRUMENT. * THE FSE INDICATED THAT THE PROBLEM WAS OBSERVED DURING AN INSTRUMENT INSTALLATION WHEN THE LASER RELAY WAS TRIPPED ACCIDENTALLY CAUSING THE ARGON LASER TO SHUTDOWN. * AS OBSERVED BY THE FSE, THE LASER TO SHUTDOWN. WAS NOT DETECTED BY THE SOFTWARE; THE FAILURE WAS OT DISPLAYED ON THE STATUS BAR AND THE SOFTWARE CONTINUED TO RUN. * THE FSE INDICATED THAT NO ONE WAS EXPOSED TO THE LASER RADATION SINCE THE LASER IS TURNED OFF BY AN INTERLOCK BREACH. * THE FSE INDICATED THAT PATIENT RESULTS WERE NOT AFFECTED IN THIS EVENT SINCE THE LASER FAILURE OCCURRED DURING THE INSTRUMENT INSTALLATION, AND NO PATIENT SAMPLES WERE TESTED AT THE TIME. * THE FSE INDICATED THAT THERE WAS NO AFFECT TO THE PATIENT OR USER AS THIS EVENT OCCURRED DURING THE INSTRUMENT INSTALLATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FC500-MPL WITH MXP SOFTWARE | GENERAL PURPOSE LABORATORY EQUIPMENT LABELED OR PROMOTED FOR | LXG | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |