CAPTURE-S
Report
- Report Number
- 1034569-2006-00066
- Event Type
- Malfunction
- Date Received
- July 24, 2006
- Date of Event
- June 30, 2006
- Report Date
- July 24, 2006
- Manufacturer
- IMMUCOR, INC.
- Product Code
- GMQ
- PMA / PMN Number
- bk950017
- Removal / Correction Number
- FA-06-06
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- OTHER
Narratives
RETENTION CAPTURE-S INDICATOR CELLS, LOT 229018 WERE TESTED ON AN IN-HOUSE SEMI-AUTOMATED INSTRUMENT WITH RETENTION CAPTURE-S PLATES, LOT S098; CAPTURE-S REACTIVE CONTROL SERUM, LOT 252007; CAPTURE-S NONREACTIVE CONTROL SERUM, LOT 251006 AND IN-HOUSE DONOR SAMPLES. IN-HOUE DONOR SAMPLES WERE NONREACTIVE, AS EXPECTED. CAPTURE-S REACTIVE CONTROL SERUM, LOT 252007 WAS TESTED AS A CONTROL AND AS TEST SAMPLES. CAPTURE-S REACTIVE CONTROL SERUM WAS INTERPRETED AS REACTIVE WHEN USED AS A CONTROL SAMPLE WITH RETENTION LOT OF CAPTURE-S INDICATOR CELLS; IT WAS INTERPRETED AS EQUIVOCAL WHEN READ AS A CONTROL SAMPLE USING THE RETURNED CAPTURE-S INDICATOR CELLS. THE CAPTURE-S REACTIVE CONTROL SERUM WAS INTERPRETED AS EQUIVOCAL IN ALL TESTING WHEN TESTED AS A TEST SAMPLE. INVESTIGATION IS ONGOING. MANUAL CAPTURE-S TESTING WAS PERFORMED. VISUAL SCORING OF CAPTURE-S REACTIVE CONTROL SERUM WAS 4 TO 6, REACTIVE. CAPTURE-S NONREACTIVE CONTROL SERUM WAS NONREACTIVE, AS EXPECTED.
CUSTOMER REPORTS THAT THEY ARE GETTING UNEXPECTED NEGATIVE REACTIONS FOR SYPHILIS TESTING WITH CAPTURE-S REACTIVE CONTROL SERUM WHEN USING CAPTURE-S INDICATOR CELLS, LOT 229018. THE FIRST TESTING RUN PERFORMED AS EXPECTED, BUT THE 2ND TESTING RUN GAVE NEGATIVE REACTIONS WITH THE REACTIVE CONTROL. THE CUSTOMER ALSO REPORTED THAT THEY WERE GETTING SOME REACTIVE SYPHILIS TEST RESULTS WITH PT SAMPLES; THESE RESULTS WERE CONFIRMED USING RPR TESTING. NO MEDICAL ACTION WAS TAKEN BASED ON THE UNEXPECTED NEGATIVE RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPTURE-S | GMQ | GMQ | IMMUCOR, INC. | * | 229018 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | CAPTURE LISS, LOT #216031.| CAPTURE-S NONREACTIVE CONTROL SERUM, LOT#251006| CAPTURE-S REACTION CONTROL SERUM, LOT 252007| CAPTURE-S MICROTITRATION WELLS, LOT S098 |