BD PRECISIONGLIDE¿ NEEDLE
Report
- Report Number
- 1911916-2018-00183
- Event Type
- Malfunction
- Date Received
- April 23, 2018
- Date of Event
- February 21, 2018
- Report Date
- April 20, 2018
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMI
- UDI-DI
- 00382903051106
- PMA / PMN Number
- K021475
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: ONE SAMPLE WAS RETURNED TO THE MANUFACTURING PLANT FOR EVALUATION. A QUALITY ENGINEER INSPECTED THE SAMPLE AND FOUND THAT THERE WAS A 19GA. 1" NEEDLE PACKAGED WITH THE 26GA 3/8" NEEDLES THUS VERIFYING THE REPORTED ISSUE. MIX-UPS CAN HAPPEN IF THERE WAS AN ISSUE WITH LINE CLEARANCE AFTER A LOT WAS PRODUCED. A DEVICE HISTORY RECORD WAS PERFORMED FOR THE REPORTED LOT AND FOUND ZERO DEFECTS NOTED FOR MIXED PRODUCT DURING PRODUCTION. INVESTIGATION CONCLUSION: QN REVIEW: NO NOTIFICATIONS WERE WRITTEN FOR THIS BATCH. DHR REVIEW: FIFTEEN VISUAL INSPECTIONS WERE PERFORMED ON 900 PARTS WITH ZERO DEFECTS NOTED FOR MIXED PRODUCT. THE BATCH PACKAGED PRIOR TO THIS ONE WAS T19 X 1 SB. INVESTIGATION RESULTS: ONE SAMPLE WAS RETURNED. THERE IS A 19 GA. 1" NEEDLE PACKAGED WITH THE 26 GA. 3/8" NEEDLES. ROOT CAUSE DESCRIPTION: IMPROPER LINE CLEARANCE. RATIONALE: THE AQL FOR MIXED PRODUCT IS (B)(4). THE SIZE OF THIS BATCH WAS (B)(4). ONE INCORRECT NEEDLE IS A DEFECT RATE OF (B)(4). SINCE THE AQL IS NOT EXCEEDED, NO CAPA WILL BE ISSUED.
IT WAS REPORTED THAT THE BD PRECISIONGLIDE¿ NEEDLE WAS MIXED IN THE INSIDE PACKING. FOUND BEFORE USE. NO OTHER SPECIFICS GIVEN. NO REPORTS OF SERIOUS INJURY OR MEDICAL INTERVENTION NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 292749 | BD PRECISIONGLIDE¿ NEEDLE | NEEDLE | FMI | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 6179848 | 00382903051106 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |