FDA Adverse Event Malfunction Summary report: N

TRUETRACK

MDR report key: 7450789 · Received April 23, 2018

Report

Report Number
1000113657-2018-00463
Event Type
Malfunction
Date Received
April 23, 2018
Date of Event
April 2, 2018
Report Date
April 23, 2018
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00041220842438
PMA / PMN Number
K032657
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(MANUFACTURER NARRATIVE = T, CORRECTED DATA = F) (B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. TEST STRIPS NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE : MLC-55 USER HAD AN INACCURATE REFERENCE: COMPETITOR'S METER: THE END USER IS COMPARING RESULTS OBTAINED FROM TRIVIDIA'S BGM SYSTEM TO THE RESULTS FROM A COMPETITOR'S BGM SYSTEM. TEST STRIP (B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER ON 4/9/2018 IN A FOLLOW-UP CALL IN ORDER TO ENSURE THAT THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN; CUSTOMER STATED THAT THE REPLACEMENT PRODUCT IS WORKING TO THEIR SATISFACTION AND THEY HAVE NOT HAD ANY MEDICAL INTERVENTION SINCE THE LAST CALL.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 223, 369, 174, 416 AND 381 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 80-150 MG/DL. CUSTOMER STATED THE TRUETRACK METER WAS READING HIGHER THAN HIS OTHER METER; ACTUAL METER TO METER COMPARISON RESULTS WERE NOT PROVIDED. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. THE CUSTOMER DECLINED TO PERFORM A BACK TO BACK BLOOD TEST DURING THE CALL ON (B)(6) 2018. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 05/31/2020 AND OPEN VIAL DATE IS (B)(6) 2018. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294471 TRUETRACK BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUETRACK RU5212 00041220842438

Patients

Seq Age Sex Outcome Treatment
1 0 YR