FDA Adverse Event Injury Summary report: N

SM HPS DV 460CC

MDR report key: 7448921 · Received April 20, 2018

Report

Report Number
1645337-2018-02259
Event Type
Injury
Date Received
April 20, 2018
Date of Event
September 5, 2017
Report Date
March 26, 2018
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001676
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ON 08/08/2018, IT WAS REPORTED TO MENTOR THAT THE DEVICE EVALUATION AND INVESTIGATION HAVE BEEN COMPLETED. UPON RECEIPT BY MENTOR, THE DEVICE CONTAINED NO FLUID. DURING EVALUATION TWO CREASES WERE OBSERVED IN THE POSTERIOR ASPECT. LEAK TESTING WAS PERFORMED IN ACCORDANCE WITH MENTOR PROCEDURES AND IT REVEALED A VALVE LEAK. NO OTHER ANOMALIES WERE DISCOVERED. DEFLATION COMPLAINT WAS CONFIRMED. MENTOR PERFORMS 100% INSPECTION AND TESTING OF ALL DEVICES PRIOR TO RELEASE. A ROOT CAUSE FAILURE ANALYSIS WILL NOT BE CONDUCTED SINCE THE INVESTIGATION OF THE REPORTED FAILURE COULD NOT CONFIRM THAT AN ACTUAL FAILURE OF THE DEVICE TO CONFORM TO EXPECTED PERFORMANCE HAD OCCURRED. THE SEVERITY LEVEL FOR THIS COMPLAINT CODE WAS DETERMINED TO BE A S3, WHICH IS DEFINED AS: NECESSITATES INTRUSIVE MEDICAL OR SURGICAL INTERVENTION. BECAUSE PRODUCT EVALUATION TEAM WAS UNABLE TO CONFIRM ANY UNUSUAL FAILURE MODE, NO CORRECTIVE ACTION WILL BE TAKEN AT THIS TIME. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A DEVICE HISTORY RECORD REVIEW IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: SALINE MENTOR SMOOTH ROUND HIGH PROFILE 460CC, CATALOG NUMBER 3503460, LOT NUMBER 231495. (B)(4).

Additional Manufacturer Narrative · 1

ON 06/02/18, IT WAS REPORTED TO MENTOR THAT THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

ON 05/04/2018, IT WAS REPORTED TO MENTOR THAT THE DEVICE HISTORY RECORD (DHR) OF LOT NUMBER 231495 REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6)-YEAR-OLD FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION PROCEDURE WITH A SALINE MENTOR SMOOTH ROUND HIGH PROFILE 460CC AND EXPERIENCED DEFLATION ON THE LEFT BREAST IMPLANT AND RIGHT CAPSULAR CONTRACTURE BAKER GRADE IV ON THE RIGHT BREAST IMPLANT DIAGNOSED BY THE HEALTHCARE PROFESSIONAL¿S EXAM. AS A RESULT, THE PATIENT HAD UNDERGONE REMOVAL AND REPLACEMENT WITH GEL MENTOR MEMORYGEL BREAST IMPLANTS 600CC. THE PATIENT HAS FULLY RECOVERED AFTER THE SURGERY. THE PATIENT HAD NO COMPLICATIONS AFTER THE SURGERY. THIS MEDWATCH IS FOR THE LEFT BREAST IMPLANT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291394 SM HPS DV 460CC PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 231495 00081317001676

Patients

Seq Age Sex Outcome Treatment
1 52 YR Required Intervention