STEM COLLARED HA COATED STD STEM SIZE 4
Report
- Report Number
- 3005180920-2018-00257
- Event Type
- Injury
- Date Received
- April 20, 2018
- Date of Event
- March 23, 2018
- Report Date
- April 20, 2018
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- UDI-DI
- 07630030804595
- PMA / PMN Number
- K121011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
BATCH REVIEW PERFORMED ON 17 APRIL 2018; LOT 172626: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 11 OCTOBER 2017. EXPIRATION DATE: 2022-10-01; NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. PRELIMINARY INVESTIGATION PERFORMED BY R&D PRODUCT MANAGER ON 19 APRIL 2018: EVALUATING THE IMAGES ATTACHED TO THIS COMPLAINT IT SEEMS THAT THE BROACH IS FUNCTIONALLY CONFORM. THERE ARE ONLY SOME LITTLE SCRATCHES VISIBLE ON THE BROACH SURFACE, DUE TO THE USAGE, THAT WE CAN CONSIDER ACCEPTABLE. THERE IS NO EVIDENCE RELATED TO THE FACT THAT THE SURGEON WASN'T ABLE TO INSERT THE STEM SIZE 4 IN THE FEMUR AFTER CANAL PREPARATION. VISUAL INSPECTION PERFORMED BY R&D PRODUCT MANAGER ON 20 APRIL 2018: THE STEM APPEARED WITHOUT ANY PARTICULAR SIGN, EXCEPT FOR THE SCRATCHES RELATED TO THE DIFFICULT REMOVAL (SIGNS UNDER THE COLLAR, GROOVE IN THE NECK). FROM THE RECEIVED PIECE IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF THE EVENT.
AFTER HAVING USED THE BROACH SIZE #4, THE SURGEON TRIED TO IMPLANT THE STEM SIZE #4, BUT IT FLOATED ABOUT 7MM UP. THE SURGEON REPLACED IT WITH A STEM SIZE #3. IT TOOK ABOUT 40 MINUTES FOR RETRIEVING. THE BROACH WAS NOT DEFECTIVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 289664 | STEM COLLARED HA COATED STD STEM SIZE 4 | CEMENTLESS HIP STEM | LZO | MEDACTA INTERNATIONAL SA | 172626 | 07630030804595 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Other |