FDA Adverse Event Injury Summary report: N

STEM COLLARED HA COATED STD STEM SIZE 4

MDR report key: 7446376 · Received April 20, 2018

Report

Report Number
3005180920-2018-00257
Event Type
Injury
Date Received
April 20, 2018
Date of Event
March 23, 2018
Report Date
April 20, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804595
PMA / PMN Number
K121011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 17 APRIL 2018; LOT 172626: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 11 OCTOBER 2017. EXPIRATION DATE: 2022-10-01; NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. PRELIMINARY INVESTIGATION PERFORMED BY R&D PRODUCT MANAGER ON 19 APRIL 2018: EVALUATING THE IMAGES ATTACHED TO THIS COMPLAINT IT SEEMS THAT THE BROACH IS FUNCTIONALLY CONFORM. THERE ARE ONLY SOME LITTLE SCRATCHES VISIBLE ON THE BROACH SURFACE, DUE TO THE USAGE, THAT WE CAN CONSIDER ACCEPTABLE. THERE IS NO EVIDENCE RELATED TO THE FACT THAT THE SURGEON WASN'T ABLE TO INSERT THE STEM SIZE 4 IN THE FEMUR AFTER CANAL PREPARATION. VISUAL INSPECTION PERFORMED BY R&D PRODUCT MANAGER ON 20 APRIL 2018: THE STEM APPEARED WITHOUT ANY PARTICULAR SIGN, EXCEPT FOR THE SCRATCHES RELATED TO THE DIFFICULT REMOVAL (SIGNS UNDER THE COLLAR, GROOVE IN THE NECK). FROM THE RECEIVED PIECE IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF THE EVENT.

Description of Event or Problem · 1

AFTER HAVING USED THE BROACH SIZE #4, THE SURGEON TRIED TO IMPLANT THE STEM SIZE #4, BUT IT FLOATED ABOUT 7MM UP. THE SURGEON REPLACED IT WITH A STEM SIZE #3. IT TOOK ABOUT 40 MINUTES FOR RETRIEVING. THE BROACH WAS NOT DEFECTIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
289664 STEM COLLARED HA COATED STD STEM SIZE 4 CEMENTLESS HIP STEM LZO MEDACTA INTERNATIONAL SA 172626 07630030804595

Patients

Seq Age Sex Outcome Treatment
1 40 YR Other