FDA Adverse Event Malfunction Summary report: N

INSERTION HANDLE F/TROCHANTERIC FIXATION NAILS

MDR report key: 7443037 · Received April 19, 2018

Report

Report Number
2939274-2018-51570
Event Type
Malfunction
Date Received
April 19, 2018
Report Date
January 11, 2018
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
FZX
UDI-DI
10886982196347
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PATIENT INFORMATION NOT AVAILABLE FOR REPORTING. DHR REVIEW WAS COMPLETED. DEVICE HISTORY RECORD (DHR) REVIEW. PART# 357.411, LOT# 6984053. MANUFACTURED BY SUPPLIER (B)(4). MANUFACTURING DATE: 14-NOV-2012. NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. CUSTOMER QUALITY CONDUCTED AN INVESTIGATION OF THE RETURNED DEVICE. THE 357.411 INSERTION HANDLE IS A REUSABLE INSTRUMENT ROUTINELY USED IN THE TITANIUM TROCHANTERIC FIXATION NAIL (TFN) SYSTEM TO AID IN FEMORAL NAIL INSERTION. THE RETURNED 5+ YEAR OLD INSERTION HANDLE WAS RECEIVED AT CUSTOMER QUALITY (CQ) WITH SIGNIFICANT POST MANUFACTURING DAMAGE. THE DEVICE HAS SEVERAL DENTS AND SCRATCHES INCLUDING HAMMER MARKS ON THE "DO NOT STRIKE" ETCH. THEREFORE THIS COMPLAINT IS CONFIRMED FOR DAMAGE. HOWEVER THE COMPLAINT FOR MISALIGNMENT WAS NOT ABLE TO BE CONFIRMED OR REPLICATED VIA FUNCTIONAL TEST BECAUSE THE OTHER DEVICES USED IN THE CONSTRUCT WERE NOT RETURNED. THE DAMAGE IN THE HOLES COULD CONTRIBUTE TO THE REPORTED COMPLAINT CONDITION. A DIMENSIONAL INSPECTION OF FEATURES RELEVANT TO THIS COMPLAINT COULD NOT BE OBTAINED AT CQ DUE TO THE POST MANUFACTURING DAMAGE. DRAWINGS MANUFACTURED THROUGH CURRENT WERE REVIEWED DURING THIS INVESTIGATION. NO PRODUCT DESIGN ISSUES OR DISCREPANCIES WERE OBSERVED. NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE MATERIAL AND RELEVANT MATERIAL PROPERTIES WERE DETERMINED TO BE CONFORMING AT THE TIME OF MANUFACTURE BASED ON REVIEW OF THE DHR. A DEFINITIVE ROOT CAUSE WAS NOT ABLE TO BE DETERMINED. IT IS LIKELY THAT THE FORMATION OF DENTS AND HAMMER MARKS WERE CAUSED BY MISPLACED HAMMER STRIKES AND ROUGH HANDLING OVER THE DEVICE'S 5+ YEAR SERVICE LIFE WHICH HAVE LED TO THIS COMPLAINT CONDITION. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

(B)(6) REPORTED AS FOLLOW: IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING THE SURGERY, INSERTION HANDLE FOR TROCHANTERIC FIXATION NAIL WAS BENT OR DAMAGED AND DID NOT ALLOW THE AIMING ARM TO ALIGN WITH THE HOLE IN THE NAIL. THERE WERE NO FRAGMENTS GENERATED, AND THERE ARE NO MISSING PIECES. THE SURGERY WAS COMPLETED USING THE DEVICE, WITH NO REPORTED PATIENT HARM. CONCOMITANT DEVICE REPORTED: TFN AIMING ARM (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) INSERTION HANDLE FOR TROCHANTERIC FIXATION NAILS. THIS IS REPORT 2 OF 2 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
285285 INSERTION HANDLE F/TROCHANTERIC FIXATION NAILS GUIDE FZX WRIGHTS LANE SYNTHES USA PRODUCTS LLC 6984053 10886982196347

Patients

Seq Age Sex Outcome Treatment
1