FDA Adverse Event Malfunction Summary report: N

TC ELECTRODE

MDR report key: 7440994 · Received April 18, 2018

Report

Report Number
3006630150-2018-01400
Event Type
Malfunction
Date Received
April 18, 2018
Date of Event
December 20, 2017
Report Date
April 18, 2018
Manufacturer
COSMAN MEDICAL
Product Code
GXI
UDI-DI
00813250013054
PMA / PMN Number
K060799
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

EXPIRATION DATE: NA /ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED/ MODEL #: CSK-TC10/ LOT #: K510/ DESCRIPTION: CSK TC ELECTRODE, 10CM TOTAL QUANTITY: 1 MODEL #: CSK-TC10 LOT #: L467 DESCRIPTION: CSK TC ELECTRODE, 10CM TOTAL QUANTITY: 1 MODEL #: TCN-10 LOT #: 011617 DESCRIPTION: NITINOL TC ELECTRODE, 100 MM TOTAL QUANTITY: 4.

Description of Event or Problem · 1

DEVICE EVALUATION WAS PERFORMED ON THE RETURNED PRODUCTS. CSK-TC10 ELECTRODES WITH LOT NUMBERS K510 AND L467 WAS FOUND TO HAVE THE HUB COMPLETELY SEPARATED FROM THE SHAFT DUE TO EXCESSIVE MECHANICAL FORCE. TCN-10 ELECTRODES WITH LOT # 011617 (TOTAL QUANTITY 4) HAD REVEALED THERE WAS DISCOLORATION (BROWN/AMBER) IN THE EPOXY WITH CHIP OUTS. THE DISCOLORED EPOXY OF ALL FOUR ELECTRODES IS UNRELATED TO THE REPORTED COMPLAINT AND IS DUE TO EXTERNAL MECHANICAL FORCE AND THERMAL STRESS ASSOCIATED WITH MULTIPLE AUTOCLAVE CYCLES. TCN-10 ELECTRODE WITH LOT # 111516 WAS FOUND TO HAVE DISCOLORED EPOXY DUE TO EXTERNAL MECHANICAL FORCE AND THERMAL STRESS ASSOCIATED WITH MULTIPLE AUTOCLAVE CYCLES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
282474 TC ELECTRODE PROBE, RADIOFREQUENCY LESION GXI COSMAN MEDICAL TCN-10 111516 00813250013054

Patients

Seq Age Sex Outcome Treatment
1